Campbell CS230 Скачать руководство пользователя страница 26

CS230 Temperature Profiler 

17 

7.3  Changing the SDI-12 Address Using 

LoggerNet

 and a 

Datalogger 

It is possible to connect multiple CS230 or other SDI-12 sensors to a single 

datalogger control port. Each temperature sensor in the CS230, or output from 

another SDI-12 device must have a unique SDI-12 address (see TABLE 

7-4

SDI-12 Addresses and Positions

 

(p. 10)

). 

The factory-set SDI-12 addresses for the CS230 start at 1 and continue until the 

last temperature sensor. The CS230 SDI-12 address is changed in software by 

issuing the 

aAb!

 command to the CS230 over the SDI-12 interface, where 

is 

the current address and 

is the new address. The current addresses of the 

individual sensors can be found by issuing the 

a!

 command. 

Campbell Scientific dataloggers (with the exception of the CR5000) support a 

method of directly interacting with SDI-12 sensors via a terminal emulator. 

This allows you to get confirmation that the change of address has worked, 

using the 

a!

 command. This can be done using a computer running 

LoggerNet

 

to issue any valid SDI-12 command through the datalogger to the CS230 as 

described in the following sections. 

7.3.1  CR1000 and CR800 Series Dataloggers 

1.

 

Connect the CS230 to the datalogger using Control Port C1 or C3 as 

described in Section 

7.1

Wiring

 

(p. 6)

. Be sure the datalogger is not running 

a program that contains the 

SDI12Recorder()

 instruction on the port used. 

2.

 

Assuming that the datalogger is configured in 

Setup

 and able to 

communicate via 

LoggerNet

, navigate to the 

Connect

 screen. Select 

Terminal Emulator

 under the 

Datalogger

 menu. The 

Terminal 

Emulator

 window will open. In the 

Select Device

 menu, located in the 

lower left-hand side of the window, select the station. 

3.

 

Click on the 

Open Terminal

 button. 

4.

 

Press the <enter> key until the datalogger responds with the CR800 

prompt. Type 

SDI12

 and select the appropriate port. 

5.

 

If the CS230 temperature sensor addresses are unknown, then conduct a 

query for each sensor’s current SDI-12 address with the 

aI!

 command. If 

no characters are typed within 12 seconds, then the mode is exited. Once a 

complete list of addresses is gathered, you will know what block of 

addresses are required to readdress the CS230. You will also be able to 

request the related metadata so that sensor locations are confirmed. Be 

sure to reference TABLE 

7-4

SDI-12 Addresses and Positions 

(p. 10)

, for a 

list of appropriate addresses.  

Содержание CS230

Страница 1: ...INSTRUCTION MANUAL CS230 Temperature Profiler Revision 4 17 C o p y r i g h t 2 0 1 6 2 0 1 7 C a m p b e l l S c i e n t i f i c I n c...

Страница 2: ......

Страница 3: ......

Страница 4: ...er Campbell Scientific s shipping address is CAMPBELL SCIENTIFIC CANADA CORP RMA _____ 14532 131 Avenue NW Edmonton Alberta T5L 4X4 Canada For all returns the client must fill out a Statement of Produ...

Страница 5: ...protection and take other appropriate safety precautions while working on or around tripods and towers Do not climb tripods or towers at any time and prohibit climbing by other persons Take reasonabl...

Страница 6: ...ounce 28 35 g 1 lb pound weight 0 454 kg Pressure 1 psi lb in2 68 95 mb Volume 1 US gallon 3 785 litres In addition part ordering numbers may vary For example the CABLE5CBL is a CSI part number and k...

Страница 7: ......

Страница 8: ...ion 7 7 2 Reading the CS230 7 7 2 1 SDI 12 Addressing 9 7 2 2 Slow Sequence Program Instructions 10 7 2 2 1 CR1000 Program to Read the Meta Data of 15 Sensors Daily 11 7 2 3 CS230 Metadata 12 7 2 3 1...

Страница 9: ...SDI 12 transparent mode on CRBasic CR800 datalogger using control port 1 and prompting for SDI 12 addresses 18 Tables 7 1 CS230 Connection to SGB3 6 7 2 SGB3 Connection to Campbell Scientific Datalogg...

Страница 10: ...unction with the SGB3 to protect against electrical surges 3 Initial Inspection Upon receipt of the CS230 inspect the packaging and contents for damage File any damage claims with the shipping company...

Страница 11: ...Accurate and stable measurements Each sensor is individually addressed and referenced to its depth Low power consumption Digital SDI 12 output Compatible with the following dataloggers CR200 X Series...

Страница 12: ...en for its resistance to temperature extremes moisture and UV degradation However this jacket will support combustion in air It is rated as slow burning when tested according to U L 94 H B and will pa...

Страница 13: ...id assembly is referenced at the top of the cross member which is the location of the first measurement point in the assembly shown in FIGURE 6 1 Internal Sensor Position 4 This information needs be a...

Страница 14: ...ient the signal and power cable of the CS230 towards the datalogger to avoid loops or strain on the cable Also use a suitable trench or conduit to protect the signal and power cable from damage Mount...

Страница 15: ...ainst electrical surges The SGB3 connects to the datalogger using the 2 ft cable shipped with the CS230 FIGURE 7 1 SGB3 3 Line Surge Protector TABLE 7 1 CS230 Connection to SGB3 Color Description CS23...

Страница 16: ...by Using low capacitance low resistance screened cable Ensuring that the power ground cable has low resistance and is connected to the same ground reference as the datalogger control ports 7 1 2 Powe...

Страница 17: ...his instruction usually takes about 700 milliseconds to execute aAb change address a to b Where in all cases a is the address of the sensor and is the command terminator These two characters are norma...

Страница 18: ...d addresses in sequence are 1 9 A Z a z no Address 0 Sending a broadcast message with the address change can correct units that have conflicting addresses aI SDI 12 Identification command X13CAMPBELLC...

Страница 19: ...reset to factory address 7 2 2 Slow Sequence Program Instructions Use of the slow sequence program instructions should be considered if the CS230 measurement will exceed the program scan interval of...

Страница 20: ...ol port C1 on the CR1000 valid port options are 1 3 5 7 Const CS230_SDI12_Port 1 Public CS230Temp NumTempSensors As Float Dim i As Long Define Data Tables DataTable Daily 1 1 DataInterval 0 1440 Min 1...

Страница 21: ...sed to identify details of the temperature probe and its individual temperature points TABLE 7 5 Meta Data Details Name Value Range Description Serial Number 0 to 65534 The serial number that is uniqu...

Страница 22: ...e MetaData_pts CS230Meta FP2 EndTable Define Subroutines ConvertNumToSDI12address Convert SDI 12 character address 0 9 A Z and a z to number value 0 61 Function ConvertNumToSDI12address address As Lon...

Страница 23: ...Uses the control port C1 on the CR1000 valid port options are 1 3 5 7 Const CS230_SDI12_Port 1 Public CS230Temp NumTempSensors As Float Dim i As Long Define Data Tables DataTable Daily 1 1 DataInterva...

Страница 24: ...The following Sample program reads a CS230 probe that has 15 temperature Sensors Individual temperatures user minimum and maximum are recorded every 5 minutes and Meta Data is collected daily Declare...

Страница 25: ...l number will be greater than 1 If CS230Meta 1 1 1 OR IfTime 0 1440 Min Then For i 1 To NumTempSensors SDI12Recorder CS230Meta i 1 CS230_SDI12_Port ConvertNumToSDI12address i R1 1 0 0 Next EndIf Read...

Страница 26: ...as described in the following sections 7 3 1 CR1000 and CR800 Series Dataloggers 1 Connect the CS230 to the datalogger using Control Port C1 or C3 as described in Section 7 1 Wiring p 6 Be sure the d...

Страница 27: ...rent address from the above step and b is the new address The temperature sensor will change its address and the datalogger will respond with the new address and then exit the SDI 12 transparent mode...

Страница 28: ...o determine the individual addresses of each temperature sensor 2 Expected sensor address has been to match another sensor address already in use a When readdressing temperature sensors you must avoid...

Страница 29: ...CS230 Temperature Profiler 20...

Страница 30: ......

Страница 31: ...ua Apinag s nbr 2018 Perdizes CEP 01258 00 S o Paulo SP BRASIL www campbellsci com br vendas campbellsci com br Campbell Scientific Canada Corp 14532 131 Avenue NW Edmonton AB T5L 4X4 CANADA www campb...

Отзывы: