Table 6-4: GET command returned data
Example
Description
0
Polling or continuous modes
2
Message format (Basic/Partial/Full)
1
Averaging period
1
Sample timing
0
Dew heater override
0
Hood heater override
0
Dirty window compensation
1
CRC checking on received commands
7.0
Sensor power down voltage
0
Enable alarm
0
Alarm high/low
10000
Alarm level
:
Delimiting character
626C
Checksum
:
Delimiting character
0x04
1
EOX
0x0D
1
Carriage return
0x0A
Line feed
1
These values are shown in hexadecimal format not ASCII.
6.4.4 Poll command
The
POLL
command requests the current visibility and/or alarm conditions from the CS140. The
output format of this command depends on how the CS140 is configured using the
SET
command or the menu interfaces.
Example of a
POLL
request with returned message:
POLL : 0 : 0 : 3 A3B:
2 0 0 60 22.9 1 1 0 0 0 0 0 0 0 0 0 0 0 0 0 5EC7
CS140 Background Luminance Sensor
22