CRVW3 3-Channel Vibrating-Wire Datalogger
20
8.1.3 Providing Temporary Power to the CRVW3
To provide temporary power to the CRVW3 while in the lab, it is
recommended that you use pn 29796 (Power Supply 24 Vdc, 5 ft Cable). You
can also use any power supply with a 16 to 28 volt DC output. Connect the
power leads to the charge terminals located on the CRVW3.
FIGURE 8-3. Power Source Connection to the CRVW3 Charge
Terminals
If your CRVW3 uses the –ALK option, you can disconnect that power source
to preserve battery life while power is applied to the charge terminals. When
using the –RC option, you can leave the rechargeable battery connected and the
charge circuit will charge the battery while the device is powered. You can also
power the device through the terminals when no battery is present.
A USB cable will also provide power to the CRVW3. However, the USB
connection alone does not provide enough power to operate the radio or make
measurements. USB power should only be used to configure the device or to
collect data. A power source other than USB should be provided for the
CRVW3 to make measurements properly and to enable radio communications.
For CRVW3 devices with the radio option, the
RX/TX
LED is solid red when
only USB power is provided to remind the user that the radio is not operating.
The
Status
LED shows a red flash when no radio is used and only USB power
is provided to the device. Measurements made while powered only with USB
will result in NAN (invalid) for measurement output.
8.1.4 Connecting to the CRVW3
Before connecting to the CRVW3 via USB, you must ensure that the USB
driver is installed as described above. After the driver is installed on a
particular PC, it should be in-place for all future connections. Connect
pn 27555 USB cable (included with the CRVW3 shipment) between the PC
and the powered CRVW3.