Section 7. Installation
281
7.7.12.2 Field Calibration Programming
Field-calibration functionality is included in a CRBasic program through either of
the following instructions:
•
FieldCal()
— the principal instruction used for non-strain gage type
sensors. For introductory purposes, use one
FieldCal()
instruction and a
unique set of
FieldCal()
variables for each sensor. For more advanced
applications, use variable arrays.
•
FieldCalStrain()
— the principal instruction used for strain gages
measuring microstrain. Use one
FieldCalStrain()
instruction and a
unique set of
FieldCalStrain()
variables for each sensor. For more
advanced applications, use variable arrays.
FieldCal()
and
FieldCalStrain()
use the following instructions:
•
LoadFieldCal()
— an optional instruction that evaluates the validity of,
and loads values from a CAL file.
•
SampleFieldCal
— an optional data-storage output instruction that
writes the latest calibration values to a data table (not to the CAL file).
FieldCal()
and
FieldCalStrain()
use the following reserved Boolean variable:
•
NewFieldCal
— a reserved Boolean variable under CR6 control used to
optionally trigger a data storage output table one time after a calibration
has succeeded.
See
CRBasic Editor Help
for operational details on CRBasic instructions.
7.7.12.3 Field Calibration Wizard Overview
The
LoggerNet
and
RTDAQ
field calibration wizards step you through the
procedure by performing the mode-variable changes and measurements
automatically. You set the sensor to known values and input those values into
the wizard.
When a program with
FieldCal()
or
FieldCalStrain()
is running, select
LoggerNet
or
RTDAQ
|
Datalogger
|
Calibration Wizard
to start the wizard. A
list of measurements used is shown.
For more information on using the calibration wizard, consult
LoggerNet
or
RTDAQ
Help.
7.7.12.4 Field Calibration Numeric Monitor Procedures
Manual field calibration through the numeric monitor (in lieu of a CR1000KD
Keyboard / Display is presented here to introduce the use and function of the
FieldCal()
and
FieldCalStrain()
instructions. This section is not a
comprehensive treatment of field-calibration topics. The most comprehensive
resource to date covering use of
FieldCal()
and
FieldCalStrain()
is
RTDAQ
Содержание CR6 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 32: ......
Страница 36: ......
Страница 38: ......
Страница 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Страница 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Страница 136: ......
Страница 251: ...Section 7 Installation 251 FIGURE 46 Running Average Frequency Response FIGURE 47 Running Average Signal Attenuation ...
Страница 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Страница 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Страница 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Страница 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Страница 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Страница 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Страница 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Страница 542: ...Section 8 Operation 542 8 11 5 PCCard Memory Card Management FIGURE 128 CR1000KD PCCard Memory Card Management ...
Страница 546: ......
Страница 549: ...Section 9 Maintenance Details 549 FIGURE 133 Separate Back Shell from Module FIGURE 134 Disconnect Battery Connector ...
Страница 552: ......
Страница 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Страница 612: ......
Страница 648: ......
Страница 650: ......
Страница 688: ......
Страница 689: ......