Section 8. Operation
419
FIGURE 100: Input Conditioning Circuit for Period Averaging
8.1.6 Reading Smart Sensors — Details
Related Topics:
• Reading Smart Sensors — Overview
(p. 77)
• Reading Smart Sensors — Details
(p. 419)
8.1.6.1 RS-232 and TTL — Details
Related Topics:
• RS-232 and TTL — Details
(p. 419)
• Serial I/O
(p. 303)
The CR3000 can receive and record most TTL (0 to 5 Vdc) and true RS-232 data
from devices such as smart sensors. See the table CR3000 Terminal Definitions
(p. 60)
for those terminals and serial ports configurable for either TTL or true RS-
232 communications. Use of the CS I/O port for true RS-232 communications
requires use of an interface device. See Hardware, Single-Connection Comms
Devices — List
(p. 611).
If additional serial inputs are required, serial input
expansion modules can be connected. See Serial I/O Modules — List
(p. 605).
Serial data are usually captured as text strings, which are then parsed (split up) as
defined in the CRBasic program.
Note When connecting serial sensors to a C terminal configured as Rx,
the sensor power consumption may increase by a few milliamps due to
voltage clamps in the CR3000. An external resistor may need to be added
in series to the Rx line to limit the current drain, although this is not
advisable at very high baud rates. See figure Circuit to Limit C Terminal
Input to 5 Volts Dc
(p. 420).
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......