Section 8. Operation
368
Offset voltages compensation applies to bridge measurements. In addition to
RevDiff and MeasOff parameters discussed in Offset Voltage Compensation
(p.
CRBasic bridge measurement instructions include the RevEx parameter that
provides the option to program a second set of measurements with the excitation
polarity reversed. Much of the offset error inherent in bridge measurements is
canceled out by setting RevDiff, MeasOff, and RevEx to True.
Measurement speed can be slowed when using RevDiff, MeasOff, and RevEx.
When more than one measurement per sensor are necessary, such as occur with
the BrHalf3W(), BrHalf4W(), and BrFull6W instructions, input and excitation
reversal are applied separately to each measurement. For example, in the four-
wire half-bridge (BrHalf4W()), when excitation is reversed, the differential
measurement of the voltage drop across the sensor is made with excitation at both
polarities and then excitation is again applied and reversed for the measurement of
the voltage drop across the fixed resistor. Further, the results of measurement
instructions (X) must be processed further to obtain the resistance value. This
processing requires additional program execution time.
Resistive-Bridge Circuits with Voltage Excitation
Resistive-Bridge Type and
Circuit Diagram
CRBasic Instruction and
Fundamental Relationship
Relational Formulas
Half-Bridge
1
CRBasic Instruction: BrHalf()
Fundamental Relationship
2
:
Three-Wire Half-Bridge
1,3
CRBasic Instruction: BrHalf3W()
Fundamental Relationship
2
:
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......