![Campbell CR3000 Micrologger Скачать руководство пользователя страница 282](http://html1.mh-extra.com/html/campbell/cr3000-micrologger/cr3000-micrologger_operators-manual_489090282.webp)
Section 7. Installation
282
Bridge Resistor Values (mΩ)
Rf
100000
Procedure
1. Build circuit
1
:
a. Use FIGURE: PT100 BrHalf3W() Three-Wire Half-Bridge Schematic
(p.
as a template.
b. For Rf, choose a 1%, 10 ppm/°C, 10000000 mΩ (10 kΩ resistor).
2. Wire circuit to datalogger:
Use FIGURE: PT100 BrHalf3W() Three-Wire Half-Bridge Schematic
(p. 281)
as the wiring diagram.
3. Calculate excitation voltage:
Use the following equation to calculate the best excitation voltage (VX) for
the measurement range of –40 to 60 °C. The equation reduces the absolute
result by 1% to allow for resistor inaccuracy:
VX
max
= VS
max
/ (RS
max
/ (Rf + RS
max
)) • 0.99
where,
VS
max
= 50 mV (maximum voltage in the
±
50
input range)
Rf = 10000000 mΩ
RS
max
= 123240 mΩ (PT00 at 60 °C)
2
so,
VX
max
= 1626420334066 mV
4. Calibrate the PT100:
If the PRT accuracy specification is good enough, and you trust it, assume
RS
0
= 100000 mΩ. Otherwise, do the following procedure:
a. Enter CRBasic EXAMPLE: PT100 BrHalf3W() Three-Wire Half-Bridge
Calibration
(p. 284)
into the CR3000. It is already programmed with the
excitation voltage from step 3.
b. Place the PRT in an ice bath (0 °C).
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......