Section 7. Installation
275
PRT Measurement Circuit Overview
Configuration
Features
Note
•
Voltage Excitation
Four-wire half-bridge (p. 277)
•
High accuracy over long leads
•
More input terminals: four per sensor
•
Slower: four differential sub
measurements per measurement
Best voltage excitation
configuration
Three-wire half-bridge (p. 281)
•
Good accuracy over long leads.
•
Fewer input terminals: two per sensor
•
Faster: two single-ended sub
measurements per measurement
Costs less to build
Four-wire full-bridge (p. 285)
•
High resolution response to change
•
More complicated to build
•
Two input terminals per sensor
•
Two differential sub measurements
per measurement
•
Best over short leads.
•
Best resolution since the
bridge balances at the
temperature-range midpoint.
•
Current excitation
Basic four-wire circuit
•
Minimizes errors due to voltage drop
with long leads.
•
Uses off-the-shelf PRTs
•
More tolerant of lower-grade wiring
•
Two input terminals per sensor
Simplest to use
Full-bridge circuit
•
Minimizes errors due to voltage drop
over long leads.
•
More tolerant of lower-grade wiring
•
Two input terminals per sensor
Best overall
7.7.17.2 General Procedure (PRT)
Following is a general procedure for using a PRT:
1.
Build circuit.
2.
Wire circuit to the CR3000.
3.
Calculate excitation voltage or current.
4.
Calibrate PRT.
5.
Measure PRT and convert output to temperature.
Several procedures follow that step you through use of common resistive-bridge
configurations to measure
a 100 Ω PRT (a.k.a, PT100). Use the following data to
help you understand the examples:
Procedure Data
•
Units used in examples: mV (millivolts), mA (milliamperes), and mΩ
(milliohms)
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......