Terminals: SE terminals 1-16
Accuracy: ±(0.01% of measu resolution), where resolution is 0.13 µs divided by the
number of cycles to be measured
Ranges:
l
Minimum signal centered around specified period average threshold.
l
Maximum signal centered around datalogger ground.
l
Maximum frequency = 1/(2 * (minimum pulse width)) for 50% duty cycle signals
Gain Code
Option
Voltage Gain
Minimum
Peak to Peak
Signal (mV)
Maximum
Peak to Peak
Signal (V)
Minimum
Pulse Width
(µs)
Maximum
Frequency
(kHz)
0
1
500
10
2.5
200
1
2.5
50
2
10
50
2
12.5
10
2
62
8
3
64
2
2
100
5
See also
10.5.4 Current-loop measurement specifications
The datalogger makes current-loop measurements by measuring across a current-sense resistor
associated with the RS-485 resistive ground terminals RG1 & RG2.
Maximum Input Voltage: ±16 V
Resistance to Ground: 101 Ω
Current Measurement Shunt Resistance: 10 Ω
Maximum Current Measurement Range: ±80 mA
Absolute Maximum Current: ±160 mA
Resolution: ≤ 20 nA
Accuracy: ±(0.1% of r 100 nA) @ -40 to 70 °C
See also
10.6 Pulse measurement specifications
Two inputs (P1-P2) individually configurable for switch closure, high-frequency pulse, or low-
level AC measurements. See also
Digital input/output specifications
(p. 171). Each terminal has its
10. Specifications
169
Содержание CR1000X
Страница 1: ...Revision 02 21 2019 Copyright 2000 2019 Campbell Scientific ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 16: ...Appendix A Glossary 176 Appendix B Index 214 Table of Contents viii ...
Страница 20: ...l Smart sensors o SDI 12 o RS 232 o Modbus o DNP3 o TCP IP o RS 485 1 Data acquisition system components 4 ...
Страница 192: ......
Страница 237: ...wiring panel 5 Appendix B Index 220 ...