CPEC300/306/310 Closed-Path Eddy-Covariance Systems
48
TABLE 6-2. Variables from
Initial Configuration
Menu
Station Variable
Default
Description
Name of variable in
Public Table (when no
CR1000KD available)
Change Press
Source
Pick
Source
EB
Used to select the barometer to
use for measurements of
ambient pressure. Set to
EB
for
EC100 enhanced barometer,
BB
for the EC100 on-board
basic barometer, or
UB
for a
user-supplied barometer.
press_source
0 =
BB
1 =
UB
2 =
EB
Reset
Sourc
FALSE
If the variable
Pick Source
has
been changed, this variable
must be set to
TRUE
to enable
the change. The program will
return
Reset Sourc
to
FALSE
once the change has been
applied.
set_press_source_flg
–1 = True
0 = False
Shadow
Correction
Pick
Crrctn
CORRTN_OFF
Used to enable the Kaimal
sonic transducer wind
shadowing correction as
described in the CSAT3B
manual.
CORRCTN_ON
enables the correction, while
CORRCTN
_
OFF
disables it.
shadow_corr
–1 = CORRCTN_ON
0 = CORRCTN_OFF
Set Crrctn
FALSE
If the variable
Pick Crrctn
has
been changed, this variable
must be set to
TRUE
to enable
the change. The program will
return
Set Crrctn
to
FALSE
once the change has been
applied.
set_shadow_corr_flg
–1 = True
0 = False
CO2 Spn Gas
400 ppm
The dry molar mixing ratio
concentration of the CO
2
span
gas. A concentration close to
ambient is recommended.
CO2_span_gas
H2O Span TDP
10 deg C
The dew point temperature of
the H
2
O span gas (i.e., the
dewpoint temperature setting of
the dewpoint generator).
Td_span_gas
Sample Flw
8 L min
-1
The set point for total flow into
the gas analyzer.
Note: If the vortex intake is
installed, a portion of this flow
will be diverted to the vortex
bypass. See the EC155 user
manual for more details.
pump_flow_set_pt
Zro/Spn Flw
1 L min
-1
Only applicable to a CPEC310
with valve module. This is the
set point for flow of zero or
valve_flow_set_pt
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...