Appendix F. CPEC300 Series Diagnostics
F-9
TABLE F-1. Summary CPEC300 diagnostic flags encoded in diag_cpec
Binary Bit
(LSB = 1)
Numeric
Value
Relevant?
Corresponding
public variable
(bit is set if
variable=false)
Status reported in
cpec_status
EC_mode
= True
EC_mode
= False
9
256
Yes
Yes
BattVolt_OK
ERROR: Battery voltage is in
deadband
or
ERROR: Battery voltage is too low
8
128
Yes
Yes
irga_OK
ERROR: IRGA problem
7
64
Yes
No
sonic_OK
ERROR: Sonic problem
6
32
Yes
No
pump_flow_OK ERROR: Pump flow problem
5
16
Yes
No
pump_tmpr_OK ERROR: Pump temperature problem
4
8
Yes
Yes
buff_depth_OK WARNING: Buffer depth too large
3
4
No
Yes
valve_flow_OK ERROR: Valve flow problem
2
2
No
Yes
valve_tmprOK
ERROR: Valve temperature problem
1
1
No
Yes
scrub_tmprOK
ERROR: Scrub module temperature
problem
none
0
NA
NA
NA
CPEC is OK
Bit 9: Battery Voltage
If
diag_cpec
is greater than 255, this indicates bit 9 of
diag_cpec
is set. To
decode other diagnostic bits, subtract 256 from
diag_cpec
and compare the
remainder to the bit values below.
Bit 9 of
diag_cpec
is equivalent to
BattVolt_OK = False
. It indicates the
power source for a CPEC300 system has dropped below the acceptable voltage
limit. This triggers a CPEC300 system to power down as much of the system
as possible to prevent a deep discharge that might damage the user’s battery.
The pump, valves, EC155 gas head, scrub module, and their associated fans,
heaters and pumps will all be powered down. The EC100 electronics and the
CSAT3A sonic head will remain powered, however. The CPEC300 will power
back up when the supply voltage reaches an acceptable level.
There are two system configuration variables;
BATT_LOWLIMIT
and
BATT_DEADBAND
that determine the shutdown and turn-on voltage (see
(p. 43)
).
BattVolt_OK
will be set to
False
if the battery voltage
BattVolt
falls below
the shutdown limit:
BattVolt
<
BATT_LOWLIMIT
Note that
BattVolt_OK
will remain
False
until the supply voltage reaches the
turn-on voltage:
BattVolt > BATT_LO BATT_DEADBAND
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...