Appendix F. CPEC300 Series Diagnostics
F-8
If
scrub_tmpr_OK
=
True,
the scrub module temperature (
scrub_tmpr)
is
within its operating range (5 to 50 °C). It is set to
False
if it is outside this
range. This check is performed continuously and Boolean variable
scrub_tmprOK
is set accordingly.
The variable
scrub_tmpr_OK
is ignored in EC mode, but if a CPEC300
system is not in EC mode, bit 1 of
diag_cpec
will be set if
scrub_tmpr_OK
=
false
, indicating there is a problem with the temperature of the scrub module.
See notes on bit 1.
F.3 CPEC300 System Diagnostic Words
A CPEC300 system has three diagnostic words that encode multiple error
conditions as binary bits. These diagnostic words are more difficult to interpret
than the status text variables, but they are more comprehensive and require
very little storage space. They are stored in the time-series output table instead
of the text-based status messages or as individual Boolean variables. The three
diagnostic words are
diag_sonic
,
diag_irga,
and
diag_cpec
. The diagnostic
words
diag_sonic
and
diag_irga
are sent directly from the EC100 and are only
partially decoded within a CPEC300 system program as needed to affect other
variables. They are stored in the time series output table for troubleshooting in
post processing. For a complete listing of the diagnostic bits encoded by these
words, see the
EC155 CO
2
and H
2
O Closed-Path Gas Analyzer Manual
.
Diagnostic word
diag_cpec
flags several conditions specific to the normal
operating range for a CPEC300 system. Some of these conditions may or may
not be relevant, depending on if a CPEC300 system is in
EC mode
. The
diagnostic word
diag_cpec
, includes only the diagnostic flags that are relevant
depending on
EC_mode
. Any value other than zero for
diag_cpec
indicates a
problem in the present mode. TABLE
lists each of the diagnostic flags,
including which mode it applies to.
To diagnose a problem when
diag_cpec
is any number other than zero, the
value is decoded according to TABLE
available in the corresponding sections below. A user unfamiliar with
converting a decimal number to binary, may find it convenient to use a
decimal-to-binary converter that can be found on the Internet. Alternately,
follow the step-by-step troubleshooting instructions as a guide through the
conversion process.
Содержание CPEC300
Страница 1: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 10 18 Copyright 2018 Campbell Scientific Inc ...
Страница 17: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 8 FIGURE 4 8 CPEC306 system ...
Страница 54: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 45 ...
Страница 55: ...CPEC300 306 310 Closed Path Eddy Covariance Systems 46 ...