background image

 

Appendix A.  Changing COM220 

Settings 

 

Changing any of the modem’s settings may result in 

communication problems.  After changing the settings, try 

the modem locally before installing it at a remote location.  

To speed problem resolution when contacting Campbell 

Scientific for support, please inform us of any modem-

settings changes that have been made. 

A.1  DIP Switch Settings 

The following COM220 settings are changed by opening or closing DIP 

switches: 

1) Communication mode, either synchronous device communication (SDC) or 

modem enable (ME) 

2) ME baud rate 

To change DIP switch settings, the outer cover must first be removed by 

unscrewing and removing the five Philips head screws on the cover. 

It is necessary to cycle power to the COM220 for changes 

to the DIP switch settings to take effect.  This is done by 

removing the 9-pin connector for about 5 s.  As with other 

devices, if you leave the power off for too short a time, you 

may get unexpected results. 

DIP Switches 1 and 2: SDC or ME 

The COM220 supports synchronous device communication (SDC) and modem 

enable (ME) communication.  Switch 1 selects between SDC and ME.  If 

switch 1 is open, ME mode is selected.  If switch 1 is closed and switch 2 is 

open (default), SDC7 is selected.  If both switch 1 and 2 are closed, SDC8 is 

selected.  SDC is not supported by non-PakBus operating systems (such as the 

CR10X, CR23X, CR7, and CR510).  Hence, to connect with these older 

dataloggers, switch 1 must be open (that is, modem enable selected).  

DIP Switches 3 and 4: ME Baud Rates 

DIP switches 3 and 4 only have effect when the COM220 is in modem enable 

mode.  If the COM220 is in ME mode, switches 3 and 4 select the baud-rate 

setting between the COM220 and the datalogger as follows: 

CAUTION 

IMPORTANT 

A-1 

Содержание COM220

Страница 1: ...INSTRUCTION MANUAL COM220 Telephone Modem Revision 3 15 C o p y r i g h t 1 9 9 7 2 0 1 5 C a m p b e l l S c i e n t i f i c I n c ...

Страница 2: ......

Страница 3: ...ducts to CSI CSI will return such Products by surface carrier prepaid within the continental United States of America To all other locations CSI will return such Products best way CIP port of entry per Incoterms 2010 This warranty shall not apply to any Products which have been subjected to modification misuse neglect improper service accidents of nature or shipping damage This warranty is in lieu...

Страница 4: ...ng container Campbell Scientific s shipping address is CAMPBELL SCIENTIFIC INC RMA _____ 815 West 1800 North Logan Utah 84321 1784 For all returns the customer must fill out a Statement of Product Cleanliness and Decontamination form and comply with the requirements specified in it The form is available from our web site at www campbellsci com repair A completed form must be either emailed to repa...

Страница 5: ...fore beginning work Wear a hardhat and eye protection and take other appropriate safety precautions while working on or around tripods and towers Do not climb tripods or towers at any time and prohibit climbing by other persons Take reasonable precautions to secure tripod and tower sites from trespassers Use only manufacturer recommended parts materials and tools Utility and Electrical You can be ...

Страница 6: ......

Страница 7: ...onnecting to a Surge Protector 8 4 Modem Settings 9 5 Troubleshooting 10 Appendices A Changing COM220 Settings A 1 A 1 DIP Switch Settings A 1 A 2 Hayes AT Commands A 2 A 3 Downloading a New Operating System to the COM220 A 6 A 4 Program Examples A 7 A 4 1 ModemCallback Example for CR1000 A 7 A 4 2 DialModem Example for CR1000 A 8 A 4 3 P97 Instruction for CR10X A 9 A 4 4 Example Programs for Data...

Страница 8: ... surge suppressor wiring 8 3 4 Side view of surge suppressor wiring 9 A 1 COM220 to computer connection A 3 A 2 Device Configuration screen A 4 A 3 Terminal Emulator screen in Device Configuration A 5 A 4 Send OS screen in Device Configuration A 7 B 1 CS I O pinout B 1 Tables 1 1 Default DIP Switch Settings 2 3 1 Dataloggers that Require Direct 12 Vdc Connection to COM220 4 A 1 Selecting ME Baud R...

Страница 9: ...aud rates 9600 to 115200 The default setting for the COM220 at shipping is SDC7 see Appendix A Changing COM220 Settings p A 1 for details The default settings ARE NOT compatible with dataloggers that have non PakBus operating systems for example CR510 CR10X CR23X CR7 CR5000 However compatibility with non PakBus dataloggers is easily achieved by changing DIP switch settings FIGURE 1 1 If you want t...

Страница 10: ...back instruction or the DialSequence and DialModem instructions to program the datalogger to originate a call to the computer For non PakBus dataloggers for example CR510 CR10X CR23X and older PakBus dataloggers for example CR10XPB CR23XPB CR510PB use Instruction 97 to program the datalogger to originate a call to the computer Refer to your datalogger manual and LoggerNet Help files for detailed d...

Страница 11: ...x 2 6 in Weight 0 16 kg 0 35 lb 3 Installation The COM220 is designed to be used with standard analog telephone lines It will not work with a digital telephone line Connection to telephone company provided coin service central office implemented systems is prohibited Connection to party line service is subject to state tariffs 3 1 Connecting to Datalogger Connect the cable from the telephone RJ11C...

Страница 12: ...aloggers that Require Direct 12 Vdc Connection to COM220 CR10 X with silver wiring panel CR10 X with black CR10 wiring panel pn 8032 21X L serial number 13 442 or lower CR500 serial number 1764 or lower CR7 CR700X serial number 2778 or lower FIGURE 3 1 CR1000 and COM220 using remote telephone line 4 ...

Страница 13: ...round terminal GND on the COM220 and to the enclosure s earth ground connection If the site does not have a grounded enclosure connect the ground wire directly to an earth ground connection The datalogger ground should also be tied to the earth ground see FIGURE 3 1 and FIGURE 3 2 The modem must be grounded for its transient protection to work CAUTION 14 AWG Ground Wire Red 12V 14 AWG Ground Wire ...

Страница 14: ...twork map and select the correct ComPort Connection for your modem Click on the PhoneBase in your network map Select your PC modem from the drop down list or add it using the Edit Modem Database dialog box Set the Maximum Baud Rate appropriate for your modem and datalogger Click on the PhoneModem in your network map Enter the COM220 s analog phone number If your datalogger is a PakBus datalogger c...

Страница 15: ...mple programs for both a router datalogger and a remote datalogger communicating by telephone to MD485 or telephone to radio 3 5 Telephone Service If you have any questions about your telephone line such as how many pieces of equipment you can connect to it the telephone company will provide this information upon request If the telephone company requests information concerning the equipment which ...

Страница 16: ...t plastic body DC breakdown at 100 V sec is 300 to 500 V DC extinguishing at 52 V 135 V and 150 V is less than 150 ms FIGURE 3 3 and FIGURE 3 4 show the wiring for connecting the surge suppressor Color coding of wires may vary The important consideration is to follow tip all the way through and ring all the way through Getting wires crossed typically does not damage anything The phone line just wi...

Страница 17: ...510 CR10X CR23X CR7 and CR5000 However compatibility with non PakBus dataloggers is easily achieved by changing DIP switch settings FIGURE 1 1 If you want to use the COM220 as a drop in replacement for the COM210 the COM220 settings should be changed to modem enable DIP switch 1 open and 9600 baud DIP switches 3 and 4 open Instructions for configuring the COM220 are described in Appendix A Changin...

Страница 18: ...n that the COM220 is tested and approved for all foreign countries CCITT is an international consultative committee that sets international communications usage standards V 32 and V 22 are CCITT standards that the COM220 meets 5 Troubleshooting When the Campbell Scientific software cannot establish a link to a remote datalogger that is connected to a COM220 check the following 1 Verify that the mo...

Страница 19: ...uter Even if a program is minimized in windows it may put a lock on a COM port Some notebook computers do not automatically activate the COM ports Verify that the COM port you are using is activated 9 Use the Low Level Log in the Campbell Scientific software LoggerNet Status Monitor Tools Log Tool View View I O Low Level Log to see connection progress while a link is being established Assuming the...

Страница 20: ...rmed by Campbell Scientific Inc or an authorized agent of Campbell Scientific Inc For assistance in installation or troubleshooting or for repairs contact Campbell Scientific Campbell Scientific Inc 815 West 1800 North Logan Utah 84321 1784 Telephone 435 227 9000 Fax 435 227 9001 Website www campbellsci com 12 ...

Страница 21: ...witch settings to take effect This is done by removing the 9 pin connector for about 5 s As with other devices if you leave the power off for too short a time you may get unexpected results DIP Switches 1 and 2 SDC or ME The COM220 supports synchronous device communication SDC and modem enable ME communication Switch 1 selects between SDC and ME If switch 1 is open ME mode is selected If switch 1 ...

Страница 22: ...in configuration mode The COM220 can be configured using an SC532 or SC532A with a 12 Vac adapter pn 15966 and Device Configuration Utility version 1 6 or newer and Hayes AT commands see below A 2 Hayes AT Commands To configure the COM220 with Hayes AT commands you will need the Campbell Scientific Device Configuration Utility version 1 6 or newer and an SC532 or SC532A device An SC12 cable must b...

Страница 23: ...RS232 cable Leave the COM220 modem disconnected for now 3 Run Device Configuration Utility version 1 6 or newer 4 Select COM220 from the Device Configuration Utility menu FIGURE A 2 5 Make sure the appropriate serial port is selected on the left panel of the Device Configuration Utility window and click on the connect button 6 Once you have clicked on the connect button connect the peripheral port...

Страница 24: ...to the COM220 7 Follow the instructions that appear in the terminal emulator screen FIGURE A 3 8 Be sure to save all new commands before exiting by entering 5 and pressing Enter 9 When finished return DIP switch 8 on the COM220 to its open position FIGURE A 2 Device Configuration screen NOTE A 4 ...

Страница 25: ...mmands For example to change the default number of rings from 1 to 3 the command is ATS0 2 the default is 1 ring as the MPU sends ATS0 1 by default to the modem module To turn the speaker off the command is ATM0 For command changes to be effective you MUST press Enter after typing the command and you MUST press 5 to save the new command before disconnecting or exiting set up COMMANDS TO CHANGE DEF...

Страница 26: ...Leave the modem disconnected for now 2 Make sure that you have the appropriate serial port selected in the left panel of the Device Configuration Utility window 3 Click the Start button 4 In the resulting file open dialogue box that appears select the file that should be sent as the modem s operating system Do not attempt to load any operating system other than that designed for the COM220 modem 5...

Страница 27: ...s The ModemCallBack instruction assumes that LoggerNet resides at the phone number and that callback is enabled in LoggerNet Setup both for the comport and for the datalogger Public PTemp batt_volt Count Result as Long Const Abort 0 False DataTable Test5 1 1 DataInterval 0 0 Sec 10 Sample 1 batt_volt FP2 Minimum 1 batt_volt FP2 0 False Maximum 1 batt_volt FP2 False False Sample 1 PTemp FP2 Minimum...

Страница 28: ... PTemp TCvalue_F dummy Public RespStrng DialSuccess AccumDialFails as LONG DataTable DC1 1 1000 DataInterval 0 0 Sec 10 Sample 1 PTemp FP2 Sample 1 TCvalue_F FP2 Sample 1 DialSuccess FP2 EndTable DialSequence 4094 Establishes a phone connection if there isn t one already DialSuccess DialModem ComSDC7 115200 5551212 Param 2 baud rate setting affects sdc bit rate Param 4 allows OS to accept any list...

Страница 29: ...s using some other modem that he knows has a different response string than those listed above A 4 3 P97 Instruction for CR10X Instruction 97 is a Program Control instruction that makes the following types of calls Datalogger calls a computer to send data Datalogger calls a voice phone using our voice synthesized modem Datalogger calls a remote datalogger to retrieve its input locations Instructio...

Страница 30: ...t Flag 8 High 4 End P95 5 Batt Voltage P10 1 1 Loc BatVolt measure battery voltage 6 Internal Temperature P17 1 2 Loc TEMP_PANL measure panel temperature 7 If time is P92 final storage every 10 seconds 1 0 Minutes Seconds into a 2 10 Interval same units as above 3 10 Set Output Flag High Flag 0 8 Real Time P77 1 111 Day Hour Minute Seconds midnight 0000 place a time stamp 9 Sample P70 1 1 Reps 2 1...

Страница 31: ...e The remote datalogger s PakBusAddress 1 In addition the RF401s or MD485s are set for CSDC8 see the RF401 or MD485 manual for directions on configuring these devices Program for CR1000 Router Router program for data callback via router w phone link to LoggerNet Phone to MD485 or RF401 w DL at Base Router PakBus Address 2 IsRouter 1 BeaconSDC8 30 to discover Remote DL which has PBA 1 Router COM220...

Страница 32: ...er Public batt_volt Result Public PTemp TCvalue_F dummy Public RespStrng DialSuccess DataTable DC1 1 1000 DataInterval 0 0 Sec 10 Sample 1 PTemp FP2 Sample 1 TCvalue_F FP2 Sample 1 DialSuccess FP2 EndTable BeginProg Scan 2 Min 3 0 PanelTemp PTemp 250 Battery Batt_volt dummy 0 TCDiff TCvalue_F 1 mV2_5C 1 TypeT PTemp True 0 250 1 8 32 SendVariables Result ComSDC7 2 4094 0000 3500 Public Callback dum...

Страница 33: ... internal line drivers for Ring and RX Data output signals 2 input Ground 3 output Ring A logic high signifies a ring signal has been detected 4 output RX Data Serial data from COM220 5 input Modem Enable A logic high internally switches power to the modem A logic low internally shuts off power to the modem 6 input Serial Device Enable A logic high disables communication with the modem without rem...

Страница 34: ...Appendix B CS I O Connection B 2 ...

Страница 35: ... is continuously powered but draws less than 2 µA The ring signal is passed on to the datalogger through an opto coupler The datalogger responds to a ring by raising the modem enable line which internally switches the regulated 5 V supply to the rest of the modem s circuitry The modem then answers and remains off hook until it loses the carrier or the datalogger lowers the modem enable line The da...

Страница 36: ...Appendix C Theory of Operation C 2 ...

Страница 37: ...t has been tested and found to comply with the limits for a Class B computing device pursuant to Subpart J of Part 15 of FCC Rules which are designed to provide reasonable protection against such interference when operated in a COMMERCIAL ENVIRONMENT Operation of this equipment in a residential area may cause interference to radio and television reception The operator must take whatever measures a...

Страница 38: ...Appendix D FCC Warning to Users of Class B Computing Devices D 2 ...

Страница 39: ...coordinated by a representative designated by the supplier Any repairs or alterations made by the user to this equipment or equipment malfunctions may give the telecommunications company cause to request the user to disconnect the equipment Users should ensure for their own protection that the electrical ground connections of the power utility telephone lines and internal metallic water pipe syste...

Страница 40: ...Appendix E IC Information E 2 ...

Страница 41: ......

Страница 42: ...om br Campbell Scientific Canada Corp CSC 14532 131 Avenue NW Edmonton AB T5L 4X4 CANADA www campbellsci ca dataloggers campbellsci ca Campbell Scientific Centro Caribe S A CSCC 300 N Cementerio Edificio Breller Santo Domingo Heredia 40305 COSTA RICA www campbellsci cc info campbellsci cc Campbell Scientific Ltd CSL Campbell Park 80 Hathern Road Shepshed Loughborough LE12 9GX UNITED KINGDOM www ca...