CNR4 Net Radiometer
12
FIGURE 7-3. The CNR4 sensor with SOLAR and TEMP cables
FIGURE 7-4. The marks on the end of the CNR4: S for SOLAR cable,
and T for TEMP cable
The four radiation outputs can be measured using differential or single-ended
inputs on the data logger. A differential voltage measurement is recommended
because it has better noise rejection than a single-ended measurement. If
differential terminals are not available, single-ended measurements can be
used. The acceptability of a single-ended measurement can be determined by
comparing the results of single-ended and differential measurements made
under the same conditions.
When differential inputs are used, jumper the low side of the input
to
⏚
to keep the signal in common mode range.
shows the data logger connections for the
SOLAR
cable, and
shows the data logger connections for the
TEMP
cable. The
cables have the white band at the pigtail end of the cable with the colo
u
r
7-6
).
NOTE