background image

1. Introduction

The BaroVUE™10 measures barometric pressure between 500 to 1100 hPa. This range is 
equivalent to a pressure measurement taken below sea level such as in a mine to over 4,570 m 
above sea level. The BaroVUE 10 is designed for use in environmental applications and is 
compatible with all Campbell Scientific data loggers.

NOTE:
This manual provides information for CRBasic data loggers only.

2. Precautions

 

l

READ AND UNDERSTAND the 

Safety

 section at the back of this manual.

 

l

Care should be taken when opening the shipping package to not damage or cut the cable 
jacket. If damage to the cable is suspected, contact Campbell Scientific.

 

l

Authorized technician shall verify that the installation and use of this product is in 
accordance to the manufacturer’s instructions, recommendations and intended use.

 

l

Although the BaroVUE 10 is rugged, it should be handled as a precision scientific 
instrument.

 

l

DO NOT blow into the pressure tube/barb. The human lungs are capable of creating a 
pressure of approximately 150 hPa above ambient, which can damage the pressure sensors 
or cause loss of calibration.

 

l

Sensor card is not a memory card. Although inserting the card into a computer slot will not 
damage the card, it will not appear as external storage or be shown in the computer device 
manager.

 

l

Do not submerge the sensor card or the motherboard in water.

 

l

Temperatures beyond the operating temperature range may damage the electronics and 
cause loss of calibration (see 

Specifications

 (p. 7)).

 

l

Unnecessary forces applied to the sensor card can cause loss of calibration.

BaroVUE™10 Digital Barometer

     

1

Содержание BaroVUE 10

Страница 1: ...04 2021 Copyright 2021 Campbell Scientific Inc...

Страница 2: ...gramming 10 7 2 2 RS 232 programming 10 7 3 Mounting 11 7 3 1 External air intake 11 7 3 2 Typical mounting 12 7 3 3 ENC100 mounting 13 7 3 4 DIN rail mounting 14 8 Operation 15 8 1 Factory settings 1...

Страница 3: ...cknowledge active command a 35 B 1 2 Send identification command al 35 B 1 3 Start verification command aV 36 B 1 4 Address query command 36 B 1 5 Change address command aAb 36 B 1 6 Start measurement...

Страница 4: ...and use of this product is in accordance to the manufacturer s instructions recommendations and intended use l Although the BaroVUE 10 is rugged it should be handled as a precision scientific instrum...

Страница 5: ...Immediately check package contents against the shipping documentation Contact Campbell Scientific about any discrepancies The following components are included with all BaroVUE 10 assemblies Table 3 1...

Страница 6: ...calibration costs The BaroVUE 10 is a fully digital sensor Electronics within the BaroVUE 10 control the measurement made by the sensor element apply the calibration coefficients average the readings...

Страница 7: ...5 QuickStart A video that describes data logger programming using Short Cut is available at www campbellsci com videos cr1000x data logger getting started program part 3 Short Cut is an easy way to p...

Страница 8: ...asure the sensor every hour or every scan The Barometric pressure and Temperature default to hPa and deg C respectively These can be changed by clicking the Barometric pressure or Temperature box and...

Страница 9: ...lick Finish and save the program Send the program to the data logger if the data logger is connected to the computer 9 If the sensor is connected to the data logger check the output of the sensor in t...

Страница 10: ...idity 0 to 97 non condensing Supply voltage 9 to 28 VDC Current consumption at 12 VDC Idle 200 A 350 A RS 232 enabled Active 5 mA Housing material Polybutylene terephthalate PBT and polycarbonate PC I...

Страница 11: ...rison to international standards NIST 7 Installation If you are programming your data logger with Short Cut skip Wiring p 8 and Programming p 10 Short Cut does this work for you See QuickStart p 4 for...

Страница 12: ...me data logger control or U terminal Each must have a unique SDI 12 address Valid addresses are 0 through 9 a through z and A through Z For CR6 and CR1000X data loggers triggering conflicts may occur...

Страница 13: ...s are provided in the following sections Downloadable example programs are available at www campbellsci com downloads barovue10 example programs 7 2 1 SDI 12 programming The SDI12Recorder instruction...

Страница 14: ...the enclosure must vent to the atmosphere not be hermetically sealed which may require the addition of a vent hole on the outer wall The preferred method is to use a two plate pressure port such as th...

Страница 15: ...pical mounting CAUTION To prevent condensation install the sensor in an environmentally protected enclosure complete with desiccant Change desiccant at regular intervals The BaroVUE 10 is typically mo...

Страница 16: ...the BaroVUE 10 in its own enclosure see FIGURE 7 4 p 14 The ENC100 is a small enclosure with dimensions of 17 x 14 x 10 cm 6 7 x 5 5 x 3 9 in It includes a compression fitting for cable entry a vent f...

Страница 17: ...mounted to a DIN Rail using the DIN Rail mounting kit purchased separately The DIN Rail mounting kit contains the following parts Table 7 4 DIN rail components Quantity Description 1 BaroVUE 10 DIN r...

Страница 18: ...n This section discusses the following 8 1 Factory settings 16 8 2 SDI 12 commands 16 8 3 Quality metric 17 8 4 Correcting pressure to sea level 18 8 5 Device Configuration Utility 18 8 6 RS 232 polle...

Страница 19: ...command follows the same pattern as an M command with the exception that it does not require the data logger to pause its operation until the values are ready Rather the data logger picks up the data...

Страница 20: ...ture 3 Quality Metric 1 mm Hg 2 C 3 N A aM4 aC4 aMC4 or aCC4 1 Pressure 2 Temperature 3 Quality Metric 1 psia 2 F 3 N A Returns the SDI 12 address 8 3 Quality metric For each pressure value the BaroVU...

Страница 21: ...level Use the following equation to find the difference dP in pressure between the sea level and the site This equation assumes U S Standard Atmosphere and dry air Wallace J M and P V Hobbes 1977 Atmo...

Страница 22: ...1 Open the main case by unclipping the latch and opening the main door FIGURE 8 1 BaroVUE 10 with main door open 2 Open Device Configuration Utility BaroVUE 10 Digital Barometer 19...

Страница 23: ...to a computer USB port See FIGURE 8 1 p 19 6 Select the Communication Port in the left panel BaroVUE10 will appear in the selection dialog NOTE It may take a few seconds for the Communication Port to...

Страница 24: ...Connect to access the Settings Editor 8 Click the Live Data tab to view real time measurements The Quality Metric Sample value should be greater than 5 see Quality metric p 17 BaroVUE 10 Digital Barom...

Страница 25: ...ferent value then click Apply 10 Click the RS 232 tab to change the RS 232 Mode Baud Rate Parity Continuous Mode Sample Period and Continuous Mode Measurement Units After changing settings click Apply...

Страница 26: ...y p 18 4 Click the Send OS tab 5 Click Start 6 Select the file It will have an a43 extension 7 Click OK The operating system will start downloading When download is complete the BaroVUE 10 will restar...

Страница 27: ...ure 2 Temperature 3 Quality Metric 1 hPa 2 C 3 not applicable N A M1 1 Pressure 2 Temperature 3 Quality Metric 1 kPa 2 C 3 N A M2 1 Pressure 2 Temperature 3 Quality Metric 1 inHg 2 F 3 N A M3 1 Pressu...

Страница 28: ...ommunication settings to match the device If the device settings are unknown use the recovery mode to reconfigure the device settings see Recovery mode p 27 5 Press enter several times until BaroVUE10...

Страница 29: ...navigate to the main menu type and press enter l To exit without saving changes navigate to the main menu type and press enter CAUTION The menu times out after 30 seconds and unsaved changes are lost...

Страница 30: ...A M1 1 Pressure 2 Temperature 3 Quality Metric 1 kPa 2 C 3 N A M2 1 Pressure 2 Temperature 3 Quality Metric 1 inHg 2 F 3 N A M3 1 Pressure 2 Temperature 3 Quality Metric 1 mmHg 2 C 3 N A M4 1 Pressur...

Страница 31: ...port 4 Open a terminal program and set the communication settings to the factory defaults Factory settings p 16 5 Continue to press enter until BaroVUE10 appears 6 Type and press enter to view the ma...

Страница 32: ...eck the voltage to the sensor If a switched 12V terminal is used temporarily connect the red wire to a 12V terminal non switched for test purposes l Verify the BaroVUE 10 SDI 12 address matches the ad...

Страница 33: ...ell Scientific for your replacement sensor card and take this newly calibrated card with you to the site of the BaroVUE 10 installation CAUTION Before handling the BaroVUE 10 wash hands and wear glove...

Страница 34: ...If the sensor card is oriented incorrectly it will not work 4 To remove the sensor card push the card in towards the socket then release Spring tension pushes the sensor card out of the socket NOTE T...

Страница 35: ...or card into the socket until it will not go any further 12 Before closing the main case double check that the card is inserted in the correct orientation Refer to FIGURE 9 1 p 30 13 Close the main ca...

Страница 36: ...for additional refinement NOTE Once the file is edited with CRBasic Editor Short Cut can no longer be used to edit the program it created 2 To add the Short Cut wiring information into the new CRBasi...

Страница 37: ...ffer Mode videos B 1 SDI 12 command basics SDI 12 commands have three components l Sensor address a a single character and the first character of the command Use the default address of zero 0 unless m...

Страница 38: ...CR LF Extended Commands aXNNN a values CR LF 1 Information on each of these commands is given in the following sections B 1 1 Acknowledge active command a The Acknowledge Active command a is used to t...

Страница 39: ...til verification information is available n the number of values to be returned when one or more subsequent D commands are issued B 1 4 Address query command Command requests the address of the connec...

Страница 40: ...o values within the next 35 seconds 0 CR LF Within 35 seconds sensor 0 indicates that it has completed the measurement by sending a service request to the data logger 0D0 The data logger immediately i...

Страница 41: ...e of data returned and the number of values a sensor returns the data logger may need to issue aD0 up to aD9 to retrieve all data A sensor may return up to 35 characters of data in response to a D com...

Страница 42: ...ransparent mode and change the SDI 12 address of an SDI 12 sensor The steps shown in Changing an SDI 12 address p 39 are used with most Campbell Scientific data loggers B 2 1 Changing an SDI 12 addres...

Страница 43: ...press Enter In this example the sensor is connected to C3 The response Entering SDI12 Terminal indicates that the sensor is ready to accept SDI 12 commands 10 To query the sensor for its current SDI...

Страница 44: ...changed from 0 to B 12 To exit SDI 12 transparent mode click Close Terminal NOTE The transparent mode for the CR6 CR3000 CR800 series CR300 series and CR1000 data loggers is similar to that shown for...

Страница 45: ...ther manufacturer s products that are resold by Campbell Scientific are warranted only to the limits extended by the original manufacturer Refer to www campbellsci com terms warranty for more informat...

Страница 46: ...ion or Repair Reference number contact your CAMPBELL SCIENTIFIC regional office Please write the issued number clearly on the outside of the shipping container and ship as directed For all returns the...

Страница 47: ...rohibit climbing by other persons Take reasonable precautions to secure tripod and tower sites from trespassers l Use only manufacturer recommended parts materials and tools Utility and Electrical l Y...

Страница 48: ...nfo campbellsci fr www campbellsci fr Germany Location Phone Email Website Bremen Germany 49 0 421 460974 0 info campbellsci de www campbellsci de India Location Phone Email Website New Delhi DL India...

Отзывы: