Actiheart User Manual 5.1.31 (Jun 2022)
15
2.3 Actiheart 5
Clicking the Actiheart 5 tabs brings a number of functions for monitoring or
maintaining the Actiheart 5 device.
The “Read Actiheart” button will automatically attempt to read from the Actiheart
connected, regardless of which section of the software the device was set up
from. It can be used as an alternative to any of the individual specialist “Read”
functions.
The
“Test Actiheart” button will open the Actiheart 5 Test window, as shown
below.
This test window is not needed in routine use, but will allow a number of
diagnostics checks to be made if requested by CamNtech technical support. At
the top of the window in the centre, the software will indicate whether an
Actiheart 5 has been detected and whether it is in normal operating mode or
“boot” mode only.
The functions available are explained below:
•
Battery:
This will send a query to the device to report on the estimated
battery level. Note that the battery is charged continuously while the
Actiheart is connected. There is no need to take any additional action to
charge the Actiheart.
•
Get Config:
This will ask the Actiheart device to report back on its current
configuration. That includes the recording mode details and recording /
log size, but not the recording metadata.
•
Get NVData:
This function will show Non-Volatile data stored on the
Actiheart, which includes both the current recording metadata (username
and start date / time) and also any permanent stored data such as serial
number and calibration values.
•
Firmware V:
The current firmware version will be reported here. Note that
the bootloader and firmware have separate version numbers, so the value
shown here will depend on whether the device has been moved into
“boot” mode or not.