1 Safety precautions
The power analyzer
CAM mobile
has been designed as a protection class I instrument in accordance with the
safety regulations IEC/EN 61010-1. To ensure the safety of the operating staff a protective earth via the power
supply connector is mandatory and the following warnings must be adhered to. Safety is not guaranteed if
warnings are not followed or if the device is operated incorrectly or handled improperly. Reference is made to
the following warnings at corresponding points in the subsequent text to assure their compliance.
1
The power supply for the CAM mobile must be provided via the IEC power connector on
the backside of the device using a three conductor IEC power cord. Under all
circumstances this cord must provide a connection to protective earth. If no connection to
protective earth is provided or if protection is disabled by using an extension cord without
protective conductor the device may be hazardous.
2
The device may only be operated by trained personnel who are capable of recognizing
contact hazards and implementing appropriate safety precautions. Hazards are present
anywhere, where voltages with a potential of higher than 50V to ground may occur.
3
When performing measurements involving contact hazards a second person should
always be present. Typically this is the case during connecting current and voltage
inputs.
4
The connection of the measurement inputs may be performed not before the device has
been connected to protective earth in accordance with warning 1.
The maximum allowed potential of current and voltage inputs is:
300V CATIII resp. 600V CATIII
5
Only use measurement leads which are in perfect condition. Especially the insulation
may not be damaged.
It must be assumed that unexpected voltages may occur at measurement objects.
Capacitors, for example, may be dangerously charged.
6
Measurements under damp ambient conditions are not allowed. The device may not be
exposed to moisture.
7
The device must be put out of operation if a safe operation, e.g. due to visible damages,
is no longer possible. To do so, you have to switch-off all connections first. Then the
device has to be sent to our factory resp. to an authorized service center.
8
Opening the device is not allowed. Live parts may be exposed. Any intervention in the
device will void the warranty. Maintenance and repair work must be carried out by our
service center only.
Device may only be disposed of in a
professional manner !