MC-III™ Panel Mount Flow Analyzer
Section 4
Enabling 4-20 mA Output
By default, the 4-20 mA output option is disabled. To enable this feature, perform the following steps:
1.
Check the “Enable” checkbox.
2.
Click on the pencil icon next to the Low Flow Rate field to launch a data-entry window (Figure 4.29) and
enter the “low” or “zero” flow rate value that will produce a 4-mA output. Adjust the decimal position,
if necessary, using the slide bar. A flow rate less than or equal to this minimum setpoint will result in an
output of 4 mA.
3.
Click on the pencil icon next to the High Flow Rate field and enter the “high” or “full-scale” flow rate
value that will produce a 20-mA output. Adjust the decimal position, if necessary, using the slide bar. A
flow rate equal to or greater than this maximum setpoint will result in an output of 20 mA.
Figure 4.29—Data-entry screen for 4-20 mA low and high values
4-20 mA Output Testing
The MC-III software allows users to test the 4-20 mA output by entering an output current value and reading
the live value generated in response.
To use the test mode, press the “Test 4-20 mA Output” button at the bottom of the 4-20 mA Output screen.
), and the 4-20 mA Output option will be temporarily
disabled.
To test the output, enter a value in the Output Current (mA) field, and press “Set Output.”
Содержание NUFLO MC-III
Страница 1: ...NUFLO TM MC III Panel Mount Flow Analyzer User Manual Manual No 2350326 01 Rev 02 ...
Страница 36: ...36 Section 3 MC III Panel Mount Flow Analyzer ...
Страница 66: ...66 Section 4 MC III Panel Mount Flow Analyzer Figure 4 30 4 20 mA output test screen ...
Страница 86: ...86 Section 5 MC III Panel Mount Flow Analyzer ...
Страница 100: ...B 2 Appendix B MC III Panel Mount Flow Analyzer ...
Страница 118: ...C 18 Appendix C MC III Panel Mount Flow Analyzer ...
Страница 119: ...C 19 MC III Panel Mount Flow Analyzer Appendix C ...
Страница 120: ...R U S S I A 603 5569 0501 ms kl c a m com ...