66
Section 4
MC-III™ EXP Flow Analyzer
4-20 mA Output
Caution:
Before performing any 4-20 mA calibration, ensure that all peripheral equipment con-
nected to the 4-20 mA current loop is either disconnected or disabled. Calibrating and
testing the 4-20 mA output feature on the MC-III EXP with the peripheral equipment in
operation may cause false alarms or erroneous operation of the peripheral device or as-
sociated equipment.
The 4-20 mA Output screen (Figure 4.28) allows users to configure the MC-III EXP to provide a 4-20 mA
output representing flow rate. Zero and full-scale values can be configured to represent any flow rate range
within the range of the flowmeter. A test mode function is also included, allowing a user to calibrate and/or
verify the output received by an end device.
Figure 4.28—4-20 mA Output screen
The low (or zero) setting is the flow rate value that will produce a 4-mA output. The high (or full-scale)
setting is the flow rate value that will produce a 20-mA output. Typically, the high value is greater than the
low value.
Flow rates in-between the minimum and maximum rate setpoints will result in an output of current between
4 mA and 20 mA according to the following calculation:
[
] [
]
MAX
MIN
OUT
CURR
MIN
I
I
I
RATE
Low
I
High
Low
−
=
×
−
+
−
Where:
I
OUT
= output current
I
MAX
= maximum current output (20 mA)
Содержание NUFLO MC-III EXP
Страница 1: ...NUFLO TM MC III EXP Flow Analyzer User Manual Manual No 9A 50165003 Rev 06 Valves Measurement ...
Страница 18: ...12 Section 1 MC III EXP Flow Analyzer ...
Страница 35: ...29 MC III EXP Flow Analyzer Section 2 ...
Страница 46: ...40 Section 3 MC III EXP Flow Analyzer ...
Страница 61: ...55 MC III EXP Flow Analyzer Section 4 Configuration Submenus cont d Pulse Output Flow Archive Event Archive ...
Страница 81: ...75 MC III EXP Flow Analyzer Section 4 ...
Страница 93: ...87 MC III EXP Flow Analyzer Section 5 Figure 5 17 Dropdown menu for exporting log data ...
Страница 94: ...88 Section 5 MC III EXP Flow Analyzer ...
Страница 105: ...99 MC III EXP Flow Analyzer Section 6 ...
Страница 114: ...A 9 MC III EXP Flow Analyzer Appendix A ...
Страница 121: ...B 7 MC III EXP Flow Analyzer Appendix B ...
Страница 123: ...C 2 Appendix C MC III EXP Flow Analyzer ...
Страница 141: ...D 18 Appendix D MC III EXP Flow Analyzer ...
Страница 143: ......