Cambridge Sensotec Rapidox 1100E Oxygen Analyser Instruction Manual
Version 3.8
14
Last printed 29/08/2012 16:01:00
using the software provided (please see section 7.4 for further details).
Alternatively, you can reset to ‘generic’ factory defaults directly from the front
panel keypad. Press ENT to enter the user setup menu and select menu option 10
“Load Defaults” and press ENT to proceed. The next screen will say “Load
Defaults ?” with the prompt to press and hold ENT for 2 seconds to confirm. To
complete this press the ENT button until the bar graph has completed and the
screen says “Defaults Loaded”. This will give you all the basic settings that are
factory set. You will now need to re-calibrate the sensor and check all the settings
are suitable.
6.16
RS232 / RS485 Port
It is possible to factory set the data communications port to either RS232 or
RS485. Once configured at the factory it is not possible to change unless the unit
is returned to Cambridge Sensotec Ltd.
Data from the Rapidox is sent on demand to the 9-way 'D' type socket (7) on the
rear panel and can be read using a simple terminal programme such as
'HyperTerminal' in Windows, or with the user's own custom software.
Alternatively a simple and convenient data-logger programme is included with the
supplied software and is described in section 8 below.
6.16.1
RS232 Protocol
Rapidox data can be read from the RS232 port at the back of the unit. The default
RS232 configuration is 115200-8-N-1, as shown in this example for COM1:
Figure 4: Rapidox default COM properties box
Alternatively, the baud rate can be reduced to 57600 or 9600 via the “Set Baud
Rate” menu item, using the front panel keypad on the Rapidox. This setting is
stored in EEPROM.