2013 Portable Spa
LTR20131000, Rev. E
Pr
eparing f
or
Your Ne
w P
or
tab
le Spa
www.calspas.com
9
120 Volt Electrical Installation
Always follow applicable local, state and federal codes
and guidelines.
Use only a dedicated electrical line with a 15 amp
breaker.
Cord-and-plug connections may not use a cord longer
than 15 feet (4.6 m) and must be plugged into a
dedicated 15 amp GFCI connection (NEC 680.42(A)
(2)). Do not use extension cords!
Always use a weatherproof-covered receptacle.
Receptacle shall be located not less than 5 feet (1.5
m) from and not exceeding 10 feet (3.0 m) from the
inside wall of the spa. (NEC 680.43(A))
Do not bury the power cord. If your cord becomes
damaged, replace it before next usage.
All 120V spas must have a GFCI. This can be either a
15 amp GFCI receptacle or a 15 amp GFCI cord and
plug kit as shown (CKIT110 - P/N ELE09700086).
Testing the GFCI
Test the GFCI plug prior to fi rst use and periodically
when the spa is powered.
Plug in the
1.
GFCI into
the power
outlet. The
i n d i c a t o r
should turn
on.
Press the TEST button.
2.
The GFCI will trip, the
indicator will turn off,
and the spa will stop
operating.
Press the RESET button.
3.
The GFCI will reset, the
indicator will turn on
again, and the spa will turn
back on.
The spa is now safe to use.
If the GFCI trips while the spa is in use, press the
RESET button. If the GFCI does not reset, unplug the
spa and call your local Cal Spas dealer for service. DO
NOT USE THE SPA!
Содержание F1239
Страница 1: ...LTR20131000 Rev E 9 18 13 ...