Enter the actual DC load current for the CURR ZERO
parameter in the MEASUREMENT CALIBRATION screen and
ALIBRATION screen.
Save this value
6.3.3 Three Phase Mode
press the ENTER key.
Save this value by pressing the SET
key.
DC Current Full-scale:
Program the output to 0 volts on the 200 range. Apply load
resistor to the output. Refer to Table 6-1. Program 160 volts.
Enter the actual output current for the CURR FS parameter in
the MEASUREMENT C
by pressing the SET key
(Series I models only).
As indicated earlier, for three-phase power systems, repeat the preceding steps for the Phase B
and C outputs. The order in which the outputs for each phase are calibrated is not important.
e current shunt as
i is in the single-phase mode as it
6.3.4 Meas
Press the PHASE key to select each output to be calibrated. Monitor the output of the
respective phase by moving the HI input of the Digital Multimeter and th
needed. The LO input should remain connected to the common LO of the sense connector.
For the MX30-3Pi/MX45-3Pi operating in single-phase mode, it will be necessary to repeat the
current measurement calibration while the MX30-3Pi/MX45-3P
maintains specific current measurement calibration coefficients for this mode. The load values
shown in Table 6-1 for MX30-1 and MX45-1 should be used on phase A only. The voltage
measurement calibration does not have to be repeated in single-phase mode as the same
coefficients for phase A are used in either phase mode.
urement Calibration Summary
The following Table is a summary of the preceding calibration steps. The value indicated by the
Ext
rrent measured by the current shunt is called I
AC
or
I
DC
.
ernal DVM is called V or V . The cu
AC
DC
TITLE
PROGRAM/LOAD PARAMETERS
PARAMETER
ADJUST TO
AC MODE
AC Vol
240 VAC, 60 Hz, no load
VOLT FS
V
AC
t Full-scale
300 VAC Range,
AC
r
CURR FS
I
AC
Cu rent Full-scale
150 VAC Range, 120 VAC, 60 Hz, full load
to 90% of max current range.
DC MODE
DC
l
(
P/N 70
Vo t Zero
03-400 only)
400 VDC Range, +2.0 VDC, no load
VOLT ZERO
V
DC
DC Volt + Full-scale
400 VDC Range, + 320 VDC, no load
VOLT FS
V
DC
DC Current Zero
(
7003-400 only)
200 VDC Range, 160 VDC, 80 ohm load
CURR ZERO
I
DC
DC Current Full-scale
200 VDC Range, 160 VDC, full load to 90%
of max. current range.
CURR FS
I
DC
Table 6-2: Measurement Calibration Table
For a multi-phase power system, repeat Paragraph 6.3 for each phase. Move the external test
equipment to the phase that is being calibrated. Refer to Figure 6-2.
While viewing the calibration screen, press the PHASE key to select the respective phase.
MX Series
163
Содержание MX30-1
Страница 4: ...User Manual Rev AJ California Instruments 4 MX Series ...
Страница 33: ...User Manual Rev AJ California Instruments MX Series 33 ilter option schematic Figure 2 12 MX output f ...
Страница 34: ...User Manual Rev AJ California Instruments Figure 2 13 MX45 Output Noise 10 KHz 1 MHz 34 MX Series ...
Страница 50: ...User Manual Rev AJ California Instruments Figure 3 11 MX135 or MX135 MB Output Wiring Rear view 50 MX Series ...
Страница 147: ...User Manual Rev AJ California Instruments Output Status Available on top assembly 7003 427 only MX Series 147 ...
Страница 148: ...User Manual Rev AJ 148 California Instruments MX Series Figure 5 2 MX Series Detailed Block Diagram ...
Страница 149: ...User Manual Rev AJ MX Series California Instruments 149 Figure 5 3 Power Module Detailed Block Diagram ...
Страница 195: ...User Manual Rev AJ California Instruments Figure 9 4 Voltage Modulation MX Series 195 ...