![California Instruments MX30-1 Скачать руководство пользователя страница 156](http://html1.mh-extra.com/html/california-instruments/mx30-1/mx30-1_user-manual_3472136156.webp)
User Manual - Rev AJ
California Instruments
6. Calibration
The Routine Calibration should be performed every 12 months. Non-routine Calibration is only
required if a related assembly is replaced or if the periodic calibration is unsuccessful.
Calibration of the MX system can be performed from the front panel or over the bus. The MXG
program provides several calibration screens for this routine calibration but not for non-periodi
calibration. This section covers calibration from the front panel. Refer to the MXGUI on line he
for information on using the MXGUI program to perform routine calibration.
UI
c
lp
C
Full-scale output calibration is done using the internal measurement system. As such, it is
important to calibrate the AC and DC voltage measurements before performing an AC and D
full-scale output calibration.
Note: Perform the Measurement calibration first.
The cardinal calibration points used during calibration are chosen to obtain optimal perform
at the typical operating points of the MX Series. If the typical application in which the MX sys
is used is unusual, it may be better to calibrate it at different operating points than the ones used
in this manual. Also, if the required load values for current calibration are not available, the
programmed voltage may be adjusted to obtain the approximate current (typically close to
maximum available current per phase).
ance
tem
6.1 Recommended Calibration Equipment
Digital Multimeter:
Fluke 8506A or equivalent / better.
1 mOhm Current Shunt:
Isotek Model RUG-Z-R001-0.1.
Load Bank:
Various high power load resistors or a resistive load bank will
be needed. (E.g. Avtron ) Size of the load bank depends on
model and phase mode. A load is required to perform the
current measurement calibration near full scale. Current
measurement calibration should be done on the lowest
available voltage range.
The accuracy and value of the load resistor is not critical as
long as the current drawn is sufficient to operate the AC Source
in the upper current range (80-100 %). Suggested values of
load bank settings are shown in Table 6-1.
PC with CI MXGUI:
Optional.
For MX30-3Pi or MX45-3Pi programmable Impedance calibration only:
Digital Phase Meter:
Krohn-Hite model 6610 or 6620 (GPIB) or equivalent.
0.02° accuracy, 0.01° resolution or better.
MX
Series
156
Содержание MX30-1
Страница 4: ...User Manual Rev AJ California Instruments 4 MX Series ...
Страница 33: ...User Manual Rev AJ California Instruments MX Series 33 ilter option schematic Figure 2 12 MX output f ...
Страница 34: ...User Manual Rev AJ California Instruments Figure 2 13 MX45 Output Noise 10 KHz 1 MHz 34 MX Series ...
Страница 50: ...User Manual Rev AJ California Instruments Figure 3 11 MX135 or MX135 MB Output Wiring Rear view 50 MX Series ...
Страница 147: ...User Manual Rev AJ California Instruments Output Status Available on top assembly 7003 427 only MX Series 147 ...
Страница 148: ...User Manual Rev AJ 148 California Instruments MX Series Figure 5 2 MX Series Detailed Block Diagram ...
Страница 149: ...User Manual Rev AJ MX Series California Instruments 149 Figure 5 3 Power Module Detailed Block Diagram ...
Страница 195: ...User Manual Rev AJ California Instruments Figure 9 4 Voltage Modulation MX Series 195 ...