User Manual – Rev AJ
California Instruments
4.7 Tr
n
ansie t Programming
4.7.1 Introduction
Transient programming provides a precise timing control over output voltage and frequency
changes. This mode of operation can be used to test a product for susceptibility to common AC
line conditions such as surges, sags, brownouts and spikes. By combining transient
programming with custom waveforms [
3Pi Controller only
], virtually any AC condition can be
simulated on the output of the AC source.
FIXED which means the output voltage is constant and remains at
the level set by the user. Changes made to the output voltage made from the PROGRAM 1
4.7.2 U
The default voltage mode is
menu take effect immediately. In front panel operation mode, the voltage and frequency slew
rates (rate of change) are always at their maximum of 1E9 V/s and 1E9 Hz/s. Slew rate
programming is only possible over the remote control interface. On power up, the AC source
always reverts to the maximum slew rate for both voltage and frequency.
sing Transient Modes
The voltage can be programmed in the following transient operating modes:
STEP
causes the output to permanently change to its triggered value.
PULSE
causes the output to change to its triggered value for a specific time, as
determined by the Pulse menu parameters.
LIST
causes the output to sequence through a number of values, as determined by
points entered in the List menu.
FIXED
disables transient operation for the selected function.
4.7.3 Step Transients
Step transients let you specify an alternate or triggered voltage level that the AC source will
apply to the output when it receives a trigger. Because the default transient voltage level is zero
volts, you must first enter a triggered voltage before you can trigger the AC source to change the
output amplitude. Step transients can only be programmed through the bus, not the front panel.
Refer to the SCPI Programming Manual for more information about programming Step
transients and triggers.
MX Series
137
Содержание MX30-1
Страница 4: ...User Manual Rev AJ California Instruments 4 MX Series ...
Страница 33: ...User Manual Rev AJ California Instruments MX Series 33 ilter option schematic Figure 2 12 MX output f ...
Страница 34: ...User Manual Rev AJ California Instruments Figure 2 13 MX45 Output Noise 10 KHz 1 MHz 34 MX Series ...
Страница 50: ...User Manual Rev AJ California Instruments Figure 3 11 MX135 or MX135 MB Output Wiring Rear view 50 MX Series ...
Страница 147: ...User Manual Rev AJ California Instruments Output Status Available on top assembly 7003 427 only MX Series 147 ...
Страница 148: ...User Manual Rev AJ 148 California Instruments MX Series Figure 5 2 MX Series Detailed Block Diagram ...
Страница 149: ...User Manual Rev AJ MX Series California Instruments 149 Figure 5 3 Power Module Detailed Block Diagram ...
Страница 195: ...User Manual Rev AJ California Instruments Figure 9 4 Voltage Modulation MX Series 195 ...