User Manual – Rev B
FCS Series II
4.8 Setting the Power-on Initialization Values
The power source is shipped with default factory settings when the unit is powered up. The
factory settings are:
Parameter
Factory default setting
Voltage 0.0
Volt
Frequency 50
Hz
Current limit
Maximum available current.
Output state
OFF
Local / Remote State
Local. Front panel unlocked.
Table 4-3: Factory Default Power on Settings
It is possible to change the power on initialization values in one of two ways:
1. Using the IEEE-488, RS232C, LAN or USB bus interface.
2. Using the front panel.
To change the power on initialization values from the front panel, proceed as follows:
1. Set the AC power source output parameters from the front panel as you want to power
up the unit.
2. Save this setting to setup register 0 from the REGISTERS menu.
3. Select the CONFIGURATION menu and move to the POWER ON field.
4. Change the POWER ON field to REG0.
5. This will recall the settings contained in register 0 at power up.
California Instruments
86
Содержание FCS Series II
Страница 4: ...User Manual Rev B FCS Series II California Instruments 4...
Страница 27: ...User Manual Rev B FCS Series II California Instruments 27...
Страница 28: ...User Manual Rev B FCS Series II Figure 3 1 Rear Panel Connector Locations California Instruments 28...
Страница 35: ...User Manual Rev B FCS Series II Figure 3 4 FCS54 3 3 Wiring Diagram Three Phase mode California Instruments 35...
Страница 36: ...User Manual Rev B FCS Series II Figure 3 5 FCS54 3 1 Wiring Diagram Single Phase mode California Instruments 36...
Страница 89: ...FCS Series II 89 Figure 5 1 FCS System Block Diagram User Manual Rev B California Instruments...
Страница 92: ...User Manual Rev B FCS Series II Figure 5 2 Power Amplifier Exploded View California Instruments 92...
Страница 95: ...User Manual Rev B FCS Series II Figure 6 1 Calibration Setup California Instruments 95...
Страница 109: ...User Manual Rev B FCS Series II Figure 7 1 Front panel location Figure 7 2 Fuse Locations California Instruments 109...
Страница 118: ...User Manual Rev B FCS Series II ure 8 4 Voltage Modulation Frequency characteristics Fig California Instruments 118...