The output transient system allows sequences of programmed voltage and or frequency
changes to be executed in a time controlled manner. Changes can be either step changes
(maximum slew rate) or ramps (specified slew rates).
The section provides some examples of programming output changes (transients). Transients
are defined as a series of numbered steps in a list. The list is executed sequentially. Each step
has a number of fields that can be set by the user:
Voltage, Voltage slew rate, Frequency, Frequency slew rate, Current, Function, Dwell
time, Trigger out, Phase.
The voltage, current and frequency settings are the same as one would do form the setup
screen using the knobs. At each step, the output will be set to the specified voltage, current
and/or frequency. The rate of change for voltage and frequency is determined by the slew rate
set. Current slew is fixed at MAX and cannot be programmed.
If the voltage is changed from 10 Vac to 20 Vac and the V slew is set to 100 V/sec, the voltage
will ramp from 10 to 20 Vac in 100 msec. ( [20 - 10] / 100 = 0.1 sec). The dwell time is the time
the output will remain at this setting. In this example, it should be set long enough to reach the
final programmed value of 20 Vac, e.g. it should be at least 0.1 sec. If not, the voltage will never
reach the final value of 20 Vac before the next step in the transient list is executed. The dwell
time may be set longer than 0.1 sec in this example. If for example the dwell time is set to 1.0
sec, the voltage will ramp from 10 Vac to 20 Vac over a 0.1 sec period and then remain at 20
Vac for 0.9 sec.
Once the dwell time set for a step in the list expires, the next step is entered (if available, if not,
execution stops and the output remains at the final values set in the last step of the list.)
Note that while there are parameters for both voltage and frequency level and slew rates, there
is only one dwell time, which applies to each step in the transient list.
Front panel entry only supports the LIST mode of operation. For Pulse and Triggered modes,
the remote control interface must be used.
When entering transient lists, each list must be entered sequentially starting with step #0. If a list
point is not yet set, the step number cannot be increased past it.
The following sample illustrates the use of transient system to program controlled output
changes.
California Instruments
83
Содержание FCS Series II
Страница 4: ...User Manual Rev B FCS Series II California Instruments 4...
Страница 27: ...User Manual Rev B FCS Series II California Instruments 27...
Страница 28: ...User Manual Rev B FCS Series II Figure 3 1 Rear Panel Connector Locations California Instruments 28...
Страница 35: ...User Manual Rev B FCS Series II Figure 3 4 FCS54 3 3 Wiring Diagram Three Phase mode California Instruments 35...
Страница 36: ...User Manual Rev B FCS Series II Figure 3 5 FCS54 3 1 Wiring Diagram Single Phase mode California Instruments 36...
Страница 89: ...FCS Series II 89 Figure 5 1 FCS System Block Diagram User Manual Rev B California Instruments...
Страница 92: ...User Manual Rev B FCS Series II Figure 5 2 Power Amplifier Exploded View California Instruments 92...
Страница 95: ...User Manual Rev B FCS Series II Figure 6 1 Calibration Setup California Instruments 95...
Страница 109: ...User Manual Rev B FCS Series II Figure 7 1 Front panel location Figure 7 2 Fuse Locations California Instruments 109...
Страница 118: ...User Manual Rev B FCS Series II ure 8 4 Voltage Modulation Frequency characteristics Fig California Instruments 118...