4.5 Measurements
Standard measurements are always available through the
Meas
key on the front panel. These
measurements are spread across multiple screens to enhance readability. Switching between
these screens can be done by successively pressing the
Meas
button on the front panel. This
will cause the screen to cycle through all available measurement screens.
asic Measurements
4.5.1 B
The following three measurement screens are available:
Parameter
MEASUREMENTS 1
VOLTAGE
AC rms voltage
CURRENT
AC rms current
FREQUENCY Frequency
POWER Real
power
MEASUREMENTS 2
VA POWER
Apparent power
VAR POWER
Reactive power
POWER FACT
Power factor
CREST FACT
Crest factor
MEASUREMENTS 3
VOLT THD
Voltage distortion
CURR THD
Current distortion
PEAK CURR
Highest AC current found
PHASE
Phase angle (relative to phase A (ø1)
Note: The V and I distortion calculations are based on H2 through H50 with the fundamental
component (H1) in the denominator.
Measurements are always running in the background. When the user selects a measurement
screen for display, the AC source first updates all the measurement parameters before
displaying the requested screen. Consequently, pressing the MEAS key may not always bring
up the selected screen immediately. There will be a perceptible delay. This will prevent the
screen from appearing with invalid or blank readouts.
4.5.2 Accuracy Considerations
Note that all measurements are AC coupled only so any DC offset will not be reported.
Any measurement system has a finite accuracy
listed in Section 2. When using the AC source
these specifications when interpreting result
pronounced as the sig
easurement range. This is
II is a high power AC source
wering low power loads,
measurement inaccuracies on rms and peak current measurements will greatly affect derived
measurements such as power, power factor and crest factor.
The measurement system on the FCS Series II uses a digital data acquisition system with a 96
Ks/sec sampling rate and 16 KHz bandwidth. This means that higher frequency components of
the measured signal are filtered out. Any contribution to the rms value of voltage and current
specification. Measurement specifications are
for measurement purposes, always consider
s. Measurement inaccuracies become more
nal being measured is at the low end of the m
particularly relevant for low current measurements. The FCS Series
optimized for providing and measuring high load currents. When po
California Instruments
80
Содержание FCS Series II
Страница 4: ...User Manual Rev B FCS Series II California Instruments 4...
Страница 27: ...User Manual Rev B FCS Series II California Instruments 27...
Страница 28: ...User Manual Rev B FCS Series II Figure 3 1 Rear Panel Connector Locations California Instruments 28...
Страница 35: ...User Manual Rev B FCS Series II Figure 3 4 FCS54 3 3 Wiring Diagram Three Phase mode California Instruments 35...
Страница 36: ...User Manual Rev B FCS Series II Figure 3 5 FCS54 3 1 Wiring Diagram Single Phase mode California Instruments 36...
Страница 89: ...FCS Series II 89 Figure 5 1 FCS System Block Diagram User Manual Rev B California Instruments...
Страница 92: ...User Manual Rev B FCS Series II Figure 5 2 Power Amplifier Exploded View California Instruments 92...
Страница 95: ...User Manual Rev B FCS Series II Figure 6 1 Calibration Setup California Instruments 95...
Страница 109: ...User Manual Rev B FCS Series II Figure 7 1 Front panel location Figure 7 2 Fuse Locations California Instruments 109...
Страница 118: ...User Manual Rev B FCS Series II ure 8 4 Voltage Modulation Frequency characteristics Fig California Instruments 118...