User and Programming Manual – Rev P
California Instruments
In the event the power source does not pass the functional test, refer to the calibration
procedure in Section 6 or call California Instrument’s customer satisfaction department for
further assistance.
High Range Load
Low Range Load
Load ON/OFF Switch Oscilloscope
AC
Output
or Distortion
Analyzer
Unit
Under
270.0V
Test
DMV
AC
Input
Model
Range
Current Load
801RP 135V 6.0A 22.5
Ω
801RP 270V 3.0A 90.0
Ω
1251RP 135V 9.2A 14.6
Ω
1251RP 270V 4.6A 58.6
Ω
Figure 3-3: Functional test setup
RP Series
October 2005
11