6.4.3 Changing the Calibration Password
The factory default calibration password is 0. You can change the password when the
AC source is in calibration mode (which requires you to enter the existing password).
Proceed as follows:
Step Action
Display
1.
Begin by pressing Shift Calibration and scrolling to the
CAL ON command.
CAL ON 0.0
2.
Enter the existing password from Entry keypad and press
Enter
3.
Press Shift Calibration and scroll to the CAL PASS
command
CAL:PASS 0
4.
Enter the new password from the keypad. You can use
any number with up to six digits and an optional decimal
point.
Table 43 : Changing the calibration password
Note: If you want the calibration function to operate without requiring any
password, change the password to 0 (zero).
6.4.4 Calibration Error Messages
Errors that can occur during calibration are shown in the following table.
Error no.
Error Meaning
401
CAL switch prevents calibration
1
402
CAL password is incorrect
403
CAL mode is not enabled
404
Computed readback CAL constants are incorrect
405
Computed programming CAL constants are incorrect
406
Incorrect sequence of CAL commands
Table 44 : IEEE-488 Calibration Error Messages
1
This is a hardware disable. See section 6.4.4.
6.4.5 Calibration Over the IEEE-488
You can calibrate the AC source by using SCPI commands within your controller
programming statements. Be sure you are familiar with calibration from the front
panel before you calibrate from a controller. Each front panel calibration command
has a corresponding SCPI command.
The SCPI calibration commands are explained in chapter 4 of the AC source
Programming Manual. Calibration error messages that can occur during IEEE-488
calibration are shown in Table 44.
Содержание 3000iL
Страница 9: ...iL Series User Manual California Instruments Rev B ix page intentionally left blank...
Страница 10: ......
Страница 37: ...iL Series User Manual California Instruments Rev B 27...
Страница 84: ...iL Series User Manual 74 Rev B California Instruments Figure 21 Troubleshooting Diagram for No Output Present...
Страница 89: ...iL Series User Manual California Instruments Rev B 79 Figure 23 AC Power Source Block Diagram...
Страница 93: ...iL Series User Manual California Instruments Rev B 83 Figure 25 AC Input Board Parts Location...
Страница 94: ...iL Series User Manual 84 Rev B California Instruments Figure 26 Mother Board Parts Location...