LIMit:VOLTage?
This command queries the maximum voltage the power source will operate at in the high
voltage range. For 2001RP-AV models configured for single voltage range, only one range
is available. In this case, this command will return the value for the only available voltage
range. The presence of a single range configuration can be derived from the limit value
returned. If the voltage limit value is less than 201 volt, the 2001RP is configured for single
voltage range operation. If a value above 200 volt is returned, the 2001RP has a high and a
low voltage range. On dual range configurations, the maximum voltage in the high range is
returned and this value must be even. The low range value is half that of the high range.
On single voltage range units, the value voltage of the only available range is returned and
this value can be odd or even.
Only the Query format of this command is available to the user.
Command Syntax
LIMit:VOLTage<NRf>
Parameters
maximum voltage available in high voltage range
[command
protected]
Query Syntax
LIMit:VOLTage?
Returned Parameters
<NR2>
98 April
2007
2001RP
Содержание 2001RP
Страница 4: ...User and Programming Manual Rev N Electrical Safety Symbols Used in This Manual 2001RP April 2007 iii...
Страница 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Страница 47: ......
Страница 48: ...User and Programming Manual Rev N Figure 5 1 AC Power System Block Diagram 2001RP April 2007 39...
Страница 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Страница 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Страница 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Страница 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...
Страница 90: ......