700M HFID
6-2-15
69
2.
Calibrating Time
: during this 10-second time, the analyzer will calculate new offset and
gain factors. The calibrating time is factory-set at 10 seconds and cannot be changed by
the operator.
3.
Verifying Time:
during this time the measured value is checked to make sure it does not
deviate outside the upper or lower limit specified by the
. The
verifying time is typically set for 10 seconds.
4.
Purge After
: the operator can set the time needed to flush any remaining calibration
gases out of the analyzer before the In Cal Status is released and the measurement status
is set.
After the Auto Calibration Timing has been set, press
to save the changes.