easy2log© RT0005 Technical Information Manual 

 

11 

Manual interface behaviour 

The  easy2log

©

  RT0005  tag  has  a  simple  manual  interface  that  includes  a  button,  a  green  LED  and  a  red  LED  that, 

depending  on  the  configuration,  permits  to  check  the  status  of  the  alarms  and  to  start/stop  the  logging  activity.  By 
default, the button can be used only to check the status of the tag but not to start/stop the logging activity. Start/stop 
function can be activated setting the Button Enable (BE) bit in the CONTROL register. 

The manual interface behaves as follows: 

Pressing the button quickly (less than 2 second) activates the check status function: 

If  the  battery  level  is  still  sufficient  for  the  logging  activity  and  no  alarms  are  active  the  green  LED  will  stay  lit  for  2 
seconds. 

If  the  battery  level  is  still  enough  for  the  logging  activity  but  there  are  active  alarms  the  red  LED  will  stay  lit  for  2 
seconds. The alarms causing the red LED to ligth are: 

Memory full (see MEMFULL bit in STATUS register description). 

ETA alarm (see ETA_ALRM bit in the STATUS register description). 

Bin counters alarm (see BIN_ALMR bit in STATUS register description). 

Mean Kinetic alarm (see MKT_ALRM bit in the STATUS register description). 

Shelf life alarm (see SHL_ALRM bit in the STATUS register description). 

If the battery has not enough charge to continue the logging activity the red or green LED will blink for 2 seconds (the 
green one in the case of no alarms, red in case of some alarms). 

 

NO ALARM 

ALARM 

Temperature Status 

Green 

Red 

 

 

NO ALARM 

ALARM 

Battery Status 

No flashing 

Flashing 

Tab. 2.1: Alarm indicator 

If the BE bit is 1 (Start/stop function activated), in addition to the behaviour previously described, the manual interface 
has the following features: 

 

The start of the logging activity can be performed in the following way: 

 

Hold down the button 

 

Wait until the green LED lit and the red LED start blinking 

 

Release  and  press  again  the  button  within  5  seconds  to  confirm  the  command  otherwise  the 

command is ignored.  

 

If the command is accepted the green LED will be lit for 15 seconds.  

 

The acquisition is started. 

And, if the SD bit is 0, also the following operation is allowed: 

 

The stop of the logging activity can be performed in the following way: 

 

Hold down the button 

 

Wait until the red LED lit and the green LED start blinking 

 

Release  and  press  again  the  button  within  5  seconds  to  confirm  the  command  otherwise  the 

command is ignored.  

 

If the command is accepted the red LED will be lit for 15 seconds.  

 

The acquisition is stopped. 

Содержание Easy2log RT0005

Страница 1: ...RT0005 UHF Semi Passive Logger Tag Technical Information Manual Revision n 02 12 03 2013...

Страница 2: ...31 Changed General Information 7 Added information on RT0005ET 7 Added information on RFID Reader Development Kits 8 Changed Ordering Options 8 Added information on Standard Gen2 memory map 18 Modifi...

Страница 3: ...rm or by any means electronic mechanical recording or otherwise without the prior written permission of CAEN RFID The information contained herein has been carefully checked and is believed to be accu...

Страница 4: ...ETIME registers RW 16 words 25 BIN_THRESHOLD registers RW 16 words 25 MKT_ACTV_ENERGY registers RW 2 words 25 MKT_THRESHOLD_TEMP register RW 25 SHL_REF_TEMP register RW 25 SHL_Q10_TEMP register RW 2 w...

Страница 5: ...stogram view 14 Fig 2 4 Cooled example time vs temperature view 15 Fig 3 1 TID memory bank structure 18 Fig 3 2 RESERVED memory bank structure 19 Fig 5 1 Temperature measurement traceability chain 35...

Страница 6: ...og RT0005 Technical Information Manual 6 1 Introduction This Chapter gives general information about easy2log RT0005 UHF Semi Passive Logger Tag It contains these topics General Information Ordering O...

Страница 7: ...al memory that can contain up to 3 958 samples The user can define up to 16 temperature ranges with independent threshold alarms for a very accurate control of the temperature excursions The tag can b...

Страница 8: ...sive tag EPC C1G2 Blistered WRT0005ETAAA RT0005ET UHF Temperature tag EPC C1G2 with external probe Development Kit WR4300PDKAAA R4300 ION Demo KIT WR4300PDKGPR R4300 GPRS ION KIT WA828BTXKITA A828BT B...

Страница 9: ...pter introduces the easy2log RT0005 UHF Semi Passive Logger Tag It contains these topics Main Features Operating states Manual interface behaviour Histogram functionality description Mean Kinetic Temp...

Страница 10: ...og RT0005 tag behaves according to the following operating states Stand by Stand by is the starting state of the tag In this state the tag respond to reader interrogations the log memory is empty and...

Страница 11: ...ee MKT_ALRM bit in the STATUS register description Shelf life alarm see SHL_ALRM bit in the STATUS register description If the battery has not enough charge to continue the logging activity the red or...

Страница 12: ...ng time is updated with the value of the BIN_SAMPLETIME X register where X is the index of the new bin Histogram bins configuration can be used also to control the storage of the samples inside the lo...

Страница 13: ...d Threshold 1 sample in Bin 3 Orange Threshold 2 samples in Bin 2 Fig 2 1 Frozen example histogram view 15 C 20 C 10 C 0 C 70 C Temp Time Orange Alarm Bin 2 5 min in orange range Red Alarm Bin 3 1 min...

Страница 14: ...histogram bins for cooled food With this configuration the tag generates an alarm if the temperature rise between 20 C and 5 C for a minimum amount of time ranging from 20 min to 29 min and 59 sec or...

Страница 15: ...easy2log RT0005 Technical Information Manual 15 20 C 5 C 0 C 70 C Temp Time Blue Alarm Bin 0 20 min in blue range Fig 2 4 Cooled example time vs temperature view...

Страница 16: ...y time it samples a new temperature The remaining Shelf Life calculation can be enabled or disabled toggling the state of the ARRE flag in the CONTROL register The remaining Shelf Life is expressed in...

Страница 17: ...ormation Manual 17 3 RT0005 Memory Description This Chapter describes the memory map of easy2log RT0005 UHF Semi Passive Logger Tag It contains these topics Address space User Memory map Register desc...

Страница 18: ...ision Total size of the TID field is 64 bits The TID memory bank contains also a 32 bit SAM field used to store the address of the sensor characteristics table as defined by the ISO18000 6REV1 draft s...

Страница 19: ...h Access Password 31 16 03h Access Password 15 0 02h Kill Password 31 16 01h Kill Password 15 0 00h Fig 3 2 RESERVED memory bank structure USER MEMORY This bank contains all the configuration register...

Страница 20: ...F BIN_THRESHOLD_15 0x42 RW 0xFFFF MKT_ACTV_ENERGY_L 0x43 RW 0x0000 MKT_ACTV_ENERGY_H 0x44 RW 0x0000 MKT_THRESHOLD_TEMP 0x45 RW 0x0000 SHL_REF_TEMP 0x46 RW 0x0000 SHL_Q10_TEMP_L 0x47 RW 0x0000 SHL_Q10_...

Страница 21: ...easy2log RT0005 Technical Information Manual 21 Register Address Op Default LOG_AREA_0 0x8A RO LOG_AREA_3956 0xFFE RO Tab 3 2 User Memory map...

Страница 22: ...register If enabled the first sample s acquisition is done SAMPLING_DELAY second after the start of the logging activity Default value is 0 RFSL RW RF_Sensitivity_Level RF_Sensitivity_Level register...

Страница 23: ...of Arrival register expressed in seconds defines the maximum time computed from shipping date required to arrive at destination If the logging is not stopped before the ETA is reached an alarm is gene...

Страница 24: ...le 1 Temperature value 0x08B Sample 1 Time low 0x08C Sample 1 Time high 0x08D Sample 2 Temperature value Time is expressed in Unix format and need 2 words for a single sample Tab 3 3 Temperature Time...

Страница 25: ...sters RW 2 words Register Address Op Default MKT_ACTV_ENERGY_L 0x43 RW 0x0000 MKT_ACTV_ENERGY_H 0x44 RW 0x0000 The value expressed in floating point 32bits IEEE 754 notation represents the Mean Kineti...

Страница 26: ..._ALRM BIN_ALRM ETA_ALRM MEMFULL BAT_MSBIT BAT_LSBIT BAT_LS MS_BIT Battery_Level Battery Level is a 2 bit register representing the remaining battery capacity Value of 3 means full capacity value 0 mea...

Страница 27: ...E_H SHL_TIME_L BIN_ALARM register RO Register Address Op Default BIN_ALARM 0x55 RO 0x0000 The register is composed by the following fields 15 2 1 0 BIN15_EN BIN02_EN BIN01_EN BIN00_EN 0 0 0 0 0 Defaul...

Страница 28: ...ault value of STOP_DATE register is 0 STOP_DATE register is set to 0 after a reset operation USER_AREA registers RW 28 words Register Address Op Default USER_AREA_0 0x6E RW USER_AREA_1 0x6F RW USER_AR...

Страница 29: ...the default value since we don t need to use it In order to activate the tag and put it in Idle state the initial date configuration register INIT_DATE shall be written with a non zero value and the...

Страница 30: ...easy2log RT0005 Technical Information Manual 30 4 RT0005 Specification This Chapter contains these topics Technical Specification Air Link Protocol Characteristic Mechanical Specification...

Страница 31: ...1 03 73 and further amended and the European Directive 2002 72 CE and further amended concerning suitability of materials intended to come into contact with foodstuffs This product is conforming to th...

Страница 32: ...be either FM0 or Miller M 4 as specified in RD3 chapter 6 3 1 3 2 Miller data encoding with M 2 and M 8 are currently not supported Tag to reader data rates the RT0005 tag supports tag to interrogator...

Страница 33: ...easy2log RT0005 Technical Information Manual 33 Mechanical Specification...

Страница 34: ...easy2log RT0005 Technical Information Manual 34 5 RT0005 Temperature Measurement Traceability...

Страница 35: ...ough the Italian Metrological Research National Institute I N RI M Istituto Nazionale di Ricerca Metrologica 4 and Italian Calibration Service SIT Servizio di Taratura in Italia 5 Fig 5 1 Temperature...

Страница 36: ...Regulatory Compliance This Chapter gives information on the easy2log RT0005 UHF Semi Passive Logger Tag Regulatory Compliance It contains these topics FCC Compliance CE Compliance Enclosure degree of...

Страница 37: ...302 208 1 V 1 2 1 2008 CEI EN 50364 2002 Reference document Test report n 29346B RD1 See CE Declaration of Conformity pag 38 for the RT0005 CE Compliance Certificate Enclosure degree of protection Re...

Страница 38: ...bmitted version to the following standards CEI EN 60950 1 2004 CEI EN 55022 2009 CEI EN 61000 6 3 2007 CEI EN 61000 6 1 2007 ETSI EN 301 489 1 V 1 8 1 2008 ETSI EN 301 489 3 V 1 4 1 2002 ETSI EN 302 2...

Отзывы: