Caen ELS MAGNET F1K5-DISS-100-15 Скачать руководство пользователя страница 12

F1K5-DISS – User’s Manual

 

12 

User shall protect output contacts by placing the FAST-PS-
1K5  &  F1K5-DISS  system  inside  a  closed  rack  or  by 
restricting the access to the back side.

 

 

 

 

Thermal Management 

 

F1K5-DISS converts electrical into thermal energy, considerable amount of heat 

is generated. This fact must be taken into consideration when installing the system. Fans 
cool the dissipation element by drawing air from front side  and exhausting it out the 
back, therefore minimum 10 cm of unrestricted air space at the front and 20 cm on the 
back of the unit must be provided. 

 
If specific application requires high power dissipation for long periods (> 800 W; 

>  10  min)  it  is  recommended  to  install  FAST-PS-1K5  and  F1K5-DISS  with  1U  of 
clearance in between to mitigate heating problems and using blanking panels to prevent 
mixing hot and cold air. Proper equipment rack ventilation shall be provided to reduce 
ambient temperature not exceed 45 °C on the front of the units. 

 

 

Do not touch the hot surfaces. The temperature of the back 
and top-right surfaces may be high when the unit is operating 
at high power.

 

 

 

 
 

 

2.4

 

F1K5-DISS connection 

 

 

Turn off the AC input power before making or changing any 
rear  panel  connection.  Ensure  that  all  connections  are 
securely  tightened  before  applying  power.  There  is  a 
potential  shock  hazard  when  using  a  power  supply  with  a 
rated output greater than 60 V

 

 
 
 

WARNING 

WARNING 

Содержание MAGNET F1K5-DISS-100-15

Страница 1: ...ISS User s Manual 1 F1K5 DISS 4 Quadrant Operation Crate for FAST PS 1K5 Digital Bipolar Power Supply Series User s Manual All Rights Reserved CAEN ELS s r l Rev 1 0 January 2021 MAGNET POWER SUPPLY S...

Страница 2: ...ual 2 This product is compliant CAEN ELS s r l in AREA Science Park S S 14 km 163 5 loc Basovizza 34149 Trieste TS Italy Mail info caenels com Web www caenels com Registered office via Vetraia 11 5504...

Страница 3: ...on Crate for FAST PS 1K5 100 A 15 V F1K5D4050030 F1K5 DISS 50 30 4 Quadrant Operation Crate for FAST PS 1K5 50 A 30 V For technical assistance please refer to the following contact CAEN ELS S R L SS14...

Страница 4: ...Rack Mounting 11 Desktop Use 11 Thermal Management 12 2 4 F1K5 DISS CONNECTION 12 Power Connection 13 Status Cable Connection 13 2 5 FAST PS 1K5 CONNECTIONS 14 2 6 FAST PS 1K5 CONFIGURATION 14 2 7 FR...

Страница 5: ...F1K5 DISS User s Manual 5 Document Revisions Revision Date Comment 0 1 September 30th 2020 Preliminary 1 0 January 11th 2021 First Release...

Страница 6: ...rating Humidity 20 to 80 RH non condensing Altitude Up to 2000 m Pollution degree 2 Overvoltage Category II Storage Temperature 10 C to 60 C Storage Humidity 5 to 90 RH non condensing The following sy...

Страница 7: ...roduct within the guarantee period if the Guarantor declares that the product is defective due to workmanship or materials and has not been caused by mishandling negligence on behalf of the User accid...

Страница 8: ...Do not use the device if it is damaged Before you use the device inspect the instrument for possible cracks or breaks before each use Do not operate the device around explosives gas vapor or dust Alw...

Страница 9: ...nuously The F1K5 DISS unit adds current voltage sinking capabilities to the system so that a FAST PS 1K5 together with a F1K5 DISS becomes a full 4 quadrant system capable not only sourcing but also s...

Страница 10: ...uring proper ventilation 3 F1K5 DISS Connections Connect the dissipation unit to main power converter 4 FAST PS 1K5 Connections Connect the power supply to the AC source 5 First switch on Switch on ch...

Страница 11: ...lation allow cooling air to reach the front panel ventilation inlets Allow minimum 10 cm of unrestricted air space at the front and 20 cm on the rear of the unit Rack Mounting The FAST PS 1K5 F1K5 DIS...

Страница 12: ...equires high power dissipation for long periods 800 W 10 min it is recommended to install FAST PS 1K5 and F1K5 DISS with 1U of clearance in between to mitigate heating problems and using blanking pane...

Страница 13: ...raight into the power plug and screw them tightly Be sure that they fit well and make a good contact Connect main power cable to DISSIPATIVE UNIT connector on FAST PS 1K5 and to FAST PS 1K5 connector...

Страница 14: ...ady enabled by default To enable or disable full current sinking capability use FAST PS 1K5 s web server interface Under CONFIG tab select INTERNAL MEMORY Set Value of ID 54 Dissipative enabled on AUX...

Страница 15: ...ent when flashing interval is short warning may be for the following reasons high power high temperature or fan malfunction When interval is long the module is not active warning indicates fan malfunc...

Страница 16: ...m the update F1K5 DISS must be connected to FAST PS 1K5 which must be switched on with its output disabled Never try to update if FAST PS 1K5 output is not disabled Connect the computer s USB port to...

Страница 17: ...ontinuos Maximum Power Dissipation 1500 W 1500 W Maximum Case Temperature Input air 45 C 110 C 110 C Maximum Air Exaust Tmperature input air 45 C 80 C 80 C Fault Temperature RED LED 100 C 100 C Maximu...

Страница 18: ...F1K5 DISS User s Manual 18 4 2 Mechanical Dimensions The mechanical dimensions of the F1K5D4100015 F1K5D4050030 units are shown in mm in Figure 4 Figure 4 F1K5D4100015 F1K5D4050030 Mechanical Drawings...

Отзывы: