Step 7: Breathing Circuit Characterization Test
To perform the Breathing Circuit Characterization test:
a. Select the Breathing Circuit Characterization Test button on the Home screen.
b. When the Breathing Circuit Characterization Test prompt appears, follow the onscreen
instructions. Ensure the breathing circuit is properly connected.
Breathing Circuit Test Prompt
c. Tap
Start
.
The CAE Air1 initiates ventilation (e.g. Pressure Support mode with default settings) on a
test lung.
d. Block the breathing circuit by inserting a plug into the patient connection port of the y-
junction.
e. Depending on the degree of occlusion, the Plimit threshold will be reached immediately
sending the ventilator into Pressure Safe mode OR the Ppeak High alarm threshold will
be reached and, after three respiratory cycles, the ventilator will go into Pressure Safe
mode.
f. In Pressure Safe mode, the Breathing Circuit Obstruction Alarm will trigger as a suspected
cause for the pressure increase event.
g. In Pressure Safe mode, the electronically controlled pressure relief valve will open;
dropping the circuit pressure to atmospheric levels. After 1 second, ventilation will
attempt to resume.
When the test is complete, the test result screen is displayed.
Содержание Air1
Страница 1: ...CAEAir1 Mechanical Ventilator User Guide...
Страница 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 80: ...2020 CAE 905K710052 v 1 1 75 Interface Overview Setup Screen PC Mode Setup Screen PS Mode Setup Screen VC Mode...
Страница 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 135: ...Appendix A Parameter Descriptions 130 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...