Appendix B - Alarms
Ventilator Hardware and Software Alarms
The following table identifies the alarm indications that may occur that are specifically related to the
ventilator hardware or software. An alarm is triggered when the equipment is not setup or
functioning as designed, or when a battery level is low or needs recharging.
Some high priority ventilator malfunctions, and the breathing circuit obstruction, trigger the Fail Safe
mode. Refer to the
Interface Overview
and
Troubleshooting
sections for more information.
Alarm
Priority
Indication
Description
Ventilator
Hardware
Malfunction
High
Visible and
audible alert
and Fail Safe
screen
Ventilator malfunction detected.
If ventilator function cannot be maintained,
Fail Safe mode is initiated.
Ventilator
Hardware
Malfunction #
High
Visible and
audible alert
Ventilator malfunction detected, but venti
-
lator settings can be safely maintained.
The malfunction code number (#) is listed in
the alarm history.
Touchscreen
Failure
High
Visible and
audible alert
Touchscreen failure detected.
Breathing Circuit
Disconnect
High
Visible and
audible alert
A disconnect is detected in the breathing
circuit connected to the patient.
Breathing Circuit
Obstruction
High
Visible and
audible alert
and Fail Safe
screen
An obstruction is detected in the breathing
circuit connected to the patient.
Fail Safe mode is initiated.
Gas Supply Failure
High
Visible and
audible alert
The oxygen supply pressure level is > 57 psi
or between 20-40 psi.
Air Temperature
High
High
Visible and
audible alert
Air temperature above 45 degrees Celsius
for 10 seconds or more.
Air Temperature
Low
Medium
Visible and
audible alert
Air temperature below 15 degrees Celsius
for 10 seconds or more.
Critical Battery
High
Visible and
audible alert
Less than 5 minutes of operating time
remaining.
Low Battery
Medium
Visible and
audible alert
Less than 10 minutes of operating time
remaining.
Battery Power
Low
Visible and
audible alert
Switching ventilator to battery power.
Software
Malfunction
High
Visible and
audible alert
Loss of communication between the UI and
MPIC.
Содержание Air1
Страница 1: ...CAEAir1 Mechanical Ventilator User Guide...
Страница 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 80: ...2020 CAE 905K710052 v 1 1 75 Interface Overview Setup Screen PC Mode Setup Screen PS Mode Setup Screen VC Mode...
Страница 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 135: ...Appendix A Parameter Descriptions 130 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...