Setup
Step 2: Connect the Trauma Fill Tank Connector to the Simulator
Locate the tan
a.
FILL
port and white
VENT
port.
Connect the
b.
FILL
(with the tan label) and
VENT
(with the white label) hoses of the
Trauma Fill Tank to the corresponding ports on the simulator
Both connections must be made for correct operation.
Step 3: Pressurize the Trauma Fill Tank and Fill the On-Board Blood
Reservoir
An integrated hand pump is used to create the pressure for the Trauma Fill Tank.
WARNING
: To prevent ejected pump assembly and/or solution from striking and injuring
you, NEVER stand with your face or body directly over the top of the tank when pumping
or loosening the pump.
To operate the pump and fill the reservoir:
Unlock the pump handle by turning counter-clockwise. (Be careful not to loosen
a.
the pump from the tank.)
Stroke the pump handle up and down from 25 to 35 times to transport 2 liters of
b.
simulated blood to the on-board blood reservoir. NEVER exceed 35 strokes while
pressurizing the tank.
Lock the pump handle back into the pump assembly by turning clockwise.
c.
Watch the Overflow Bottle located on the tank assembly. When liquid begins to
d.
appear in this bottle, the on-board blood reservoir is full. (Filling the on-board
blood reservoir takes approximately 3 to 5 minutes.)