![Cadex C8000 Скачать руководство пользователя страница 85](http://html1.mh-extra.com/html/cadex/c8000/c8000_user-manual_3250378085.webp)
C8000 Battery Testing System User Manual
73
2.12
External Devices
The C8000 system supports the following third-party devices VIA
the serial ports.
2.12.1.1
Programmable DC Electronic Loads
All programmable loads are supported on Serial Port Three only.
Connect an electronic load
Configure the external load to communicate with the following settings;
BAUD: 384000 (Recommended), Data bits: 8, Parity: N, Stop Bits: 1
Connect a 9-pin Serial Cable to the C8000 and DC load.
Reboot the C8000.
Note:
1)
Ensure that all waveforms used in conjunction with an electronic load are within
the capabilities of the DC load.
2)
Some DC loads have only a TTL output. Contact the manufacturer of the DC
load for a TTL to Serial convertor.
Supported DC Electronic Loads
BK Precision 8510 - BK Precision 8514 - BK Precision 8518 - BK Precision 8520
BK Precision DC Electronic Load Limits
;
The programming of the DC Load will take between 10~90 seconds to complete. All waveforms
will be programmed at the start of the EXTERNAL LOAD steps.
Minimal pulse width supported by the BK DC Loads is 1ms.
2.12.1.2
Thermal Chambers
All programmable Thermal Chambers are supported on Serial Port Two only.
Connect to a Thermal Chamber
Configure the Thermal Chamber to communicate with the following settings;
BAUD: 19200 (Recommended), Data bits: 8, Parity: N, Stop Bits: 1
Connect a 9-pin Serial Cable to the C8000 and DC load..
Reboot the C8000.
Note:
Ensure that all program steps and commands used in conjunction, with the thermal
chambers, are within the capabilities of the thermal chamber units.
Supported Thermal Chambers
Thermotron 2800 - Thermotron 3800- Thermotron 8800- TestEquity Watlow F4