Follow sterile procedures.

1.

 

Turn unit on. “__ __” is displayed for five seconds and then 
disappears. The device is now ready for use. If “__ __” is 
displayed continuously, see Troubleshooting. 

2.

 

Remove contents of disposable pouch. 

3.

 

Place needle firmly on tapered chamber stem. 
 

4.

  Remove cap on pre-filled syringe and screw the syringe onto 

remaining chamber stem. Be careful not to contaminate the 
fluid pathway. 
 
 
 
 
 
 
 

5.

 

Open cover of monitor and inspect transducer port for damage 
or contamination. If damaged, do not use and return to C2DX. 
Remove all contamination with a soft damp cloth. Place 
chamber in well (black surface down) and gently push until 
disposable seats. 
 
 
 
 
 
 
 
 
 
 
 
 

6.

  Snap cover closed.

 DO NOT FORCE.

 

Make sure latch has 

snapped.  

NOTE: If the cover will not close, rotate the syrings so that a straight 
side of the flange rests against the base of the monitor. This will 
reduce the height of the flange to allow the cover to close.

QUICK PRESSURE MONITOR SET INSTRUCTIONS

7.

 

Hold the needle approximately 45° up from horizontal. Slowly 
force fluid through the disposable to purge it of air. 

CAUTION: DO NOT allow saline to roll down into the transducer 
well. 
 

8.

 

Approximate the intended angle of the insertion of the needle 
into the skin. To insure accuracy the display must read -8 to 
+9mm/Hg. Press the “ZERO” button. The display will read “00” 
after approximately 2 seconds.

 

 

 

NOTE: Display must read “00” before continuing. If it does not, 
please see troubleshooting.

 

9.

 

Insert needle into body. Slowly inject less than 3/10 cc of saline 
into the compartment to equillibrate with interstitial fluids.  
 

 

Wait for the display to reach equillibrium. 

 

Read pressure. 

 

For additional measurements, repeat steps 8-11. Make sure unit 
is reset to zero.

WARNING: Device must be at correct angle of 
insertion before the “ZERO” button is pressed 
to insure accuracy.

WARNING: DO NOT use pressure readings when 
display indicates low batter. Replace battery 
and repeat steps to obtain pressure reading.

STIC Intra-Compartmental Pressure  
Monitor System 

Instruction Guide

555 East Eliza Street 

Schoolcraft, MI 49087 

P: 888.902.2239

F: 269.204.2480 

c2dxinc.com