ENGLISH

haze-gard 

plus

  

Operation

On the left Side:

  Main Switch 

on/off

  Voltage Selector

  Fuse

  Connection for:

  Power Cable

  Foot Switch

  PC

How to Calibrate?

Calibration is necessary at the first 
startup and at a regular interval 
(two months recommended):

  Press set up button to call up  

  the calibration routine

  Cover haze-port with dark  

  standard and press operate

  Remove cover from haze-port  

  and start the 100%  
  measuremen

  Place the clarity standard at the  

  clarity port and press operate

operate button:

Measure

set up button:

SET-UP MENU

reference button:

Measure reference in Continuous 
mode

Cursor keys:

Used in SET-UP mode to select 
individual menu items and set 
values.
In any measuring mode, these keys 
are used to switch statistics  
functions on/off, and to delete 
values from statistics memory

SET UP MENUE 

¢

 Calibrate 

£

 Print Statistic 

£

 Print Online: 

off 

£

 Cont. Mode: 

off 

£

 Language  

USA

Escape:   

SET UP

delete all

delete last

statistic

result

on/off

on/off

operate

set up reference

trans-

mittance

haze

clarity

delete all

delete last

statistic

result

on/off

on/off

H

    1 4 . 7

    clarity-port

    haze-port

Calibration

Zero Calibration

Calibration

100% Calibration

Calibration

Place clarity stand

75.5

at clarity-port