GasAlertQuattro
User Manual
58
• Battery status
• Configuration changes
• Low and high alarm setpoints (all sensors)
• STEL and TWA alarm setpoints (H
2
S and CO only)
• STEL and TWA interval settings (H
2
S and CO only)
• Fleet Manager II options that are defined for the detector and
configuration changes
• Low alarm acknowledges
• Cal due and bump due disregarded via button press
• Calibration and bump test interval settings
• Calibration data: zero, span, time, date, and pass/fail events
• Bump test (time, date, and pass/fail events)
• Battery status
• Log type
• Temperature
• Datalog interval
• Language
• Sensor options
Bump and Calibration Results
The detector records the MicroDock II bump test and calibration results.
The results can then be imported into Fleet Manager II to create detailed
reports.
Note
To record calibration and bump test results, the detector must
be calibrated and bump tested using the MicroDock II station.
The following information is recorded in the calibration and bump test
results:
• Date/time bump tests and calibrations are performed
• Serial number
• Test performed
• Test results
• Detector status
• Detector configuration updates during bump test/calibration
• Type of detector
• MicroDock II serial number
• Last calibration performed
• Next calibration due date