GasAlertQuattro
User Manual
20
Sensors Due for Bump Test
If a bump test is due, the following screen displays.
flashes
for each sensor that is due for a bump test.
Note
N/A displays when the bump test interval has been defined as
0 days. Refer to
.
a
Caution
BW recommends to bump test the sensors before each
day’s use to confirm their ability to respond to gas by
exposing the sensors to a gas concentration that
exceeds the alarm setpoints.
Press
C
to acknowledge that a bump test is due. If
Force
Bump
is disabled, the detector enters normal operation.
Force Bump Test Enabled:
If the
Force Bump
option is
enabled, the following screen displays.
The sensor(s) must pass the bump test to enter normal opera-
tion.
Apply gas manually or via the MicroDock II station, or press
C
to exit and deactivate the detector. Refer to
Note
BW Technologies by Honeywell recommends to bump test the
sensors before each day’s use to confirm their ability to
respond to gas by exposing the sensors to a gas concentration
that exceeds the alarm setpoints. Manually verify that the audi-
ble and visual alarms activate. Calibrate if the readings are not
within the specified limits.
If error screens display, refer to