GasAlertQuattro
Activating/Deactivating the Detector
Activating/Deactivating the Detector
a
Caution
Only activate the detector in a safe area that is free of
hazardous gas in an atmosphere of 20.9% oxygen.
Activate:
Press and hold
C
.
Deactivate:
Press and hold
C
during the powering off countdown.
Release
C
when
OFF
displays.
Startup Sequence
If an error screen displays during the startup sequence, refer to
When the detector is activated, it performs several tests during the
startup sequence. Confirm the following tests occur.
Battery Test
If battery power is critically low upon startup, the detector displays the
following screens and then deactivates.
Recharge the battery for 6 hours. Refer to
Segment Test
1. The segment test verifies that the visual, audio, and
vibrate functions are operating correctly. The detector
alarms, vibrates, and displays the following screen.
Product Identification and Firmware Revision
2. The following two screens display showing the BW and product
identification, and the firmware revision.