GasAlertMicro 5 IR 

Maintenance

 

Replacing the Batteries 

a

 Warning: Use only the GasAlertMicro 5 IR battery pack with the GasAlertMicro 5 IR detector (not compatible with 

other Micro 5/PID models and vice versa). Refer to the warning label located on the battery pack lid. 

a

 Warning: Always deactivate the detector before removing the battery pack.  

The battery packs are user-changeable in hazardous areas, but the alkaline batteries inside the pack must be changed 
in a safe area that is free of hazardous gas. 

 

1. 

Open the latch on the bottom of the detector and remove the battery pack by lifting the bottom end of the pack upwards. 

2. 

Unscrew the two captive screws on the battery pack cover, open the pack, and replace the three alkaline batteries. 

3. 

Replace the battery pack cover and reinsert the screws. 

4. 

Replace the battery pack on the detector and secure the latch. 

13

 

Содержание GasAlertMicro 5 IR

Страница 1: ...GasAlertMicro 5 IR O2 CO H2S PH3 SO2 Cl2 NH3 NO2 HCN ClO2 O3 VOC CO2 and Combustibles 1 2 3 4 and 5 Gas Detectors...

Страница 2: ...dations of BW b the buyer promptly notifying BW of any defect and if required promptly making the product available for correction No goods shall be returned to BW until receipt by the buyer of shippi...

Страница 3: ...man and French The screens for the additional languages are displayed on the detector and in the corresponding Quick Reference Guides Safety Information Read First Use the detector only as specified i...

Страница 4: ...s such as leaded gasoline and halogenated hydrocarbons may temporarily inhibit sensor performance in most cases the sensor will recover after calibration For use only in potentially explosive atmosphe...

Страница 5: ...5 IR Parts of the GasAlertMicro 5 IR Item Description 1 Visual alarm bars LEDs 2 Toxic 1 IR CO2 sensor 3 Toxic 2 sensor 4 LEL sensor 5 Audible alarms 6 Oxygen sensor 7 Pushbuttons 8 Liquid crystal di...

Страница 6: ...or 3 Gas cylinder 4 Gas identifier bars 5 Battery life indicator 6 Pass code lock 7 Data transmission 8 Clock 9 Stealth mode 10 Optional pump indicator 11 Optional datalogger card indicator 12 Alarm c...

Страница 7: ...lear the TWA STEL and maximum MAX gas exposure readings press C and G simultaneously and hold until the countdown is complete To view the date and time alarm setpoints TWA STEL low and high of all the...

Страница 8: ...oes all of the sensors except CO2 and calibrates the oxygen sensor If a sensor fails to auto zero it will bypass the span 5 Select which sensor to span Press G and H to scroll to the required sensor a...

Страница 9: ...be changed for that sensor The LCD displays the following options Press C to set or press A to bypass the alarm setpoints 9 Press H or G to change the alarm setpoint Press C to save the value and pro...

Страница 10: ...Display Low Alarm Fast beep Slow flash L and target gas bar flash Vibrator alarm activates TWA Alarm Fast beep Slow flash L and target gas bar flash Vibrator alarm activates High Alarm Constant beep F...

Страница 11: ...Over Limit OL Exposure Alarm Fast beep and flash L and target gas bar flash Vibrator alarm activates Sensor Alarm One beep every 15 seconds FAIL flashes above the failed sensor Automatic Shutdown Alar...

Страница 12: ...r press C to run a pump test Two fast beeps and alternating flashes Vibrator alarm activates L and J icons flash MMC Fail Alarm One beep every 5 seconds S icon flashes Note Alarms can be set to be lat...

Страница 13: ...r that is overdue for calibration Latch Latched alarm Enable to ensure an alarm persists until it is acknowledged press C to acknowledge Passcode Passcode protection Enable to prevent unauthorized per...

Страница 14: ...ly applicable to LEL sensors Autocal Automatic oxygen calibration Enable disable the detector to automatically calibrate the oxygen sensor during start up 4 Logger Datalogging Sets how often the detec...

Страница 15: ...tery pack The battery packs are user changeable in hazardous areas but the alkaline batteries inside the pack must be changed in a safe area that is free of hazardous gas 1 Open the latch on the botto...

Страница 16: ...nt dimensions 14 5 x 7 4 x 3 8 cm 5 7 x 2 9 x 1 5 in Weight 370 g 13 1 oz Operating and storage conditions Temperature 20 C to 50 C 4 F to 122 F Humidity CO2 IR 0 to 95 relative humidity non condensin...

Страница 17: ...larm multi gas alarm over range alarm sensor alarm pump alarm MMC fail alarm low battery alarm confidence beep automatic shutdown alarm Audible alarm 95 dB at 1 ft 0 3 m variable pulsed dual beepers V...

Страница 18: ...E 0539 g II 2 G EEx d ia IIC KEMA 06ATEX 0206X Approved Class I Division 1 Group A B C and D Standards CAN CSA C22 2 No 157 and C22 2 152 ANSI UL 913 and ANSI ISA S12 13 Part 1 IECEx Ex d ia IIC This...

Страница 19: ...on Printing line...

Отзывы: