GasAlertClip Extreme English Instruction Sheet
D2139/5 Page 2
Trimmed and folded dimensions: 4.125 (w) x 5.875 (h)
Activating the Detector
Press and hold
C
for 5 seconds.
Note: Once activated the detector cannot be deactivated, except after a
battery life-ending alarm. Refer to
.
The self-test must be performed in an atmosphere that is free of
background gas.
After a self-test is performed, wait 30 seconds before using to ensure
the detector accurately detects gas.
Prior to each days use, a self-test of the detector must be performed. When
Test
displays on the LCD, a self-test is required. When the self-test is
performed and passes,
u
displays to confirm the tests are successful.
Confirm that the following tests (1-5) are performed.
Press
C
and release (approximately 1 second) to initiate the self-test. The
tests are as follows:
1. The detector emits one beep, one flash, and one vibration.
2. The detector sets the battery life-remaining clock.
3. All LCD elements display.
4.
Test
flashes while the sensor integrity and battery life are tested.
5. The low and high alarm setpoints display.
Self-Test Pass
When the self-test is successful, the detector emits a short beep and a
single vibration.
u
then displays to verify that the self-test has passed.
Twenty hours after performing the self-test,
Test
again displays to indicate
that a self-test is required.
Self-Test Fail
If the self-test fails, the detector emits five short beeps and flashes before
displaying a blank screen. The LCD then returns to the normal operating
screen and again displays
Test
.
Repeat the self-test.
Note: If the self-test fails three consecutive times, the LCD displays a blank
screen and the detector deactivates (safety shutdown mode).
Automatic Battery Test
The battery is automatically tested every 2 hours. If the battery test fails,
another automatic test is initiated 30 minutes later.
Note: After five consecutive battery test failures, the LCD displays a blank
screen and the detector deactivates (safety shutdown mode).
Detector Life-Remaining Clock
The detector life-remaining clock indicates how much longer the detector
will operate. The LCD displays the countdown value of remaining months,
days, and then hours.
The detector continues to operate for a maximum of 8 hours after the
detector life-ending alarm initiates. Press
C
to deactivate the detector. For
more information, refer to
.
Safety Shutdown Mode
The LCD displays a blank screen when initiating safety shutdown mode.
The detector beeps, flashes, and vibrates rapidly (twice per second) until
the battery loses power. To deactivate the detector before the battery loses
power, press
C
. Contact
.
The detector initiates safety shutdown mode if the
•
self-test fails three consecutive times,
•
automatic battery test fails five consecutive times, or
•
detector has not been manually deactivated within 8 hours of initiating
the detector life-ending alarm.
Gas Alarm Setpoints
The following is a list of the factory alarm setpoints.
Model
Low Alarm
Setpoint
High Alarm
Setpoint
GasAlertClip Extreme O
2
19.5%
23.5%
GasAlertClip Extreme CO
35 ppm
200 ppm
GasAlertClip Extreme H
2
S
10 ppm
15 ppm
GasAlertClip Extreme SO
2
5 ppm
10 ppm
Note: Detector may be configured with customer specified alarm setpoints.
Displaying the Gas Alarm Setpoints
Press
C
to display the alarm setpoints.
Alarm setpoints are factory configured and cannot be modified.