background image

 

 
 

 

 
 
 
 
 

 

NOVA LED

 

 
 

 
 
 

 
 

 
 
 

 
 

Floor Stand, Wall, Table and Ceiling

 

 

Instructions for Use & Maintenance

 

 

 
 
 
 

5000781 Rev A01 Instructions for Use 

Содержание NOVA LED NXC

Страница 1: ...NOVA LED Floor Stand Wall Table and Ceiling Instructions for Use Maintenance 5000781 Rev A01 Instructions for Use...

Страница 2: ...2...

Страница 3: ...3 NOVA LED Task Examination Light BASE MODEL MOUNT AREA NX T table FL floor W wall C ceiling BLANK United States 120V Example NXFL is a floor model for the US...

Страница 4: ...mbly 9 1 3 Table Mount Assembly 9 1 4 Floor Stand Assembly 9 1 5 Floor Stand Operation 10 1 6 Ceiling Assembly 10 2 Maintenance 11 2 1 LED Replacement 11 2 2 Preventive Maintenance Checklists 12 2 3 C...

Страница 5: ...g a NEMA 5 15 hospital grade North American plug An On Off On switch operates the light at the base of the light head The light is UL listed and made in the USA Specifications General The light is int...

Страница 6: ...to 75 keep dry Atmospheric pressure 500hPa to 1060hPa Shipping Container Markings Fragile Keep Dry Operation Ambient temperature 10 C to 40 C Relative humidity 30 to 75 Atmospheric pressure 700hPa to...

Страница 7: ...7 Dimensions FLOOR STAND TABLE WALL CEILING...

Страница 8: ...8 Wiring Diagram Floor Wall Table Ceiling Mount 120V 60Hz Input...

Страница 9: ...rier within five days after receipt of the shipment 2 If damage or shortage occurs in transit the delivering carrier is required by law to make notation of a shortage or damage This notation is to be...

Страница 10: ...the table clamp 1 4 Floor Stand Assembly See instructions included with Floor Stand 1 5 Floor stand Operation 1 5 1 Plug the power cord into a grounded convenience socket Note Turn the light On by pr...

Страница 11: ...hown If mounted to wood use 10 steel wood screws min 2 long Mounting to a steel support use 1 4 20 steel bolts or threaded rod with nut and lock washer backing Max gap between support and ceiling plat...

Страница 12: ...he power cord for external abrasion Lightly pull and twist the cord at the inlet strain relief to test for internal con ductor breakage when the light is energized Check electrical connectors for evid...

Страница 13: ...ore starting any repair 3 2 2 Move the light head up or down by grasping the handle 3 2 3 Do not force the light head past its stops 3 3 Warnings 3 3 1 To avoid the risk of electrical shock this equip...

Страница 14: ...termine failure mode and replace transformer Replace switch Poor low Light Output Loose or corroded connectors Faulty LED driver Check all connections clean replace as necessary Replace power supply R...

Страница 15: ...15...

Страница 16: ...16 This product was designed and assembled in the U S A by Burton Medical LLC 2300 W Windsor Ct STE C Addison IL 60101 U S A Tel 800 444 9909 Fax 800 765 1770 Web www burtonmedical com...