209
of 216
8.6 External actuation of a statistics reset
You can use the "IN_RESET" control signal to reset the part counter and NOK counter and also the
evaluation element statistics. The DIGIFORCE
®
9311 resets the statistics for all the measurement
programs in this case.
The following values are reset by this function:
•
Part counter and NOK counter for all measurement programs
•
NOK percentage for all measurement programs
Diagram 103:
External actuation of a statistics reset
Cycle
a. The PLC first checks whether the instrument is ready, i.e. if "OUT_READY" = 1.
b. The PLC initiates the reset process with "IN_RESET" = 1.
c. The DIGIFORCE
®
9311 then resets "OUT_READY" to 0.
d. On detecting "OUT_READY" = 0, the PLC withdraws the reset request with
"IN_RESET" = 0.
e. At the end of the cycle, the DIGIFORCE
®
9311 sets "OUT_READY" to 1 (in standby).
Содержание DIGIFORCE 9311