176
of 216
6.3.9 USB flash
In the "USB flash" menu (M80), you can enable/disable data logging on an external USB stick.
Note:
If you wish to log the measurement data on the USB stick, you need to individually enable data
logging for each measurement program.
1
Plug a USB stick into the rear USB port.
2
In measurement mode, tap anywhere on the touchscreen. The icon appears in the
bottom
-
right corner.
3
Tap to open the "Configuration Main Menu".
4
Tap the "Program Setup" icon.
5
Tap the "USB flash" icon.
6
Enable the checkbox under "Logging".
7
Tap to return to the "Program Setup" menu.
Note:
Make sure you use a suitable USB stick. Please refer to the data sheet for further details. The
latest version of the data sheet is available at
You can find further information in section
"USB flash" on page 57.
Содержание DIGIFORCE 9311