Digital indicator
Model 9163
Page
91
9.1 Configuring an alarm limit
Before you can configure an alarm limit, you must first enable it. This is done in the menu
Hrd
.
¾
First switch from the main menu into the menu
Pas
.
To do this, hold down the [F] button until
PAS
is displayed.
PV
¾
Use the
▲
or
▼
buttons to set the value "
99
".
¾
Then hold down the [F] button until
HRD
is displayed.
PV
¾
Press the [F] button briefly several times until
AL
.
n
is displayed.
PV
¾
Use the
▲
or
▼
buttons to set the number of alarm limits to be enabled.
You can enable up to four alarm limits.
¾
Press the [F] button to confirm the value.
This enables a specific number of alarm limits (up to four allowed). All other settings are made in the
menu
ALL
.
¾
Now go to the menu
ALL
.
To do this, hold down the [F] button until
ALL
is displayed.
PV
¾
Press the [F] button briefly.
You are now at the parameter
Ar
.
n
("
n
" stands for the number of the limit: AL1, AL2 etc.).
PV
¾
Use the code to specify the reference variable for the respective limit.