2.2 Installation / panel-mounting
2.2.1 Installation dimensions
108
96
115
44,5
92
70
48
10
170
Figure 1: Installation dimensions of the 9163 digital indicator (dimensions in millimeter)
2.2.2 Installation instructions
¾
Read the basic installation rules before you install the unit.
If you do not follow our safety instructions, there may be problems with electrical safety and
electromagnetic compatibility.
Furthermore, disregarding the safety instructions will forfeit the warranty.
Basic installation rules for the 9163
If you connect the unit to devices that are
not
electrically isolated (e.g. thermocouples):
¾
Provide a dedicated conductor for the ground connection.
The ground connection must never be made directly via the machine rack.
If you install the unit in applications in which there is a risk of personal injury, damage to machinery or
property:
¾
Combine the unit with additional limit monitoring equipment.
¾
Check regularly during operation whether the alarm limit has actuated.