5
Cleaning and Disinfection
5.1
Safety Information on Cleaning and Disinfection
CAUTION
Cleaning is the most important measure and requirement for a successful chemical disinfec-
tion.
When the bed is occupied by the same resident, routine cleaning of the bed is generally suffi-
cient.
Disinfection of the chassis is only necessary when it has been visibly contaminated with in-
fectious or potentially infectious materials (blood, stool, pus) or, under doctor’s orders, in the
presence of an infectious disease.
Before a new resident occupies the bed, it must first be cleaned and disinfected by wiping!
Failure to follow these instructions could result in considerable damage to the
bed frame and its electrical equipment and lead to subsequent faults!
•
Unplug the power plug and store it so that it does not come into excessive contact with
water or other cleaning solutions (place in a plastic bag).
•
Ensure that all plugs on the bed itself are inserted correctly in the control unit and the
drive motors.
•
Ensure that none of the electrical components show any signs of external damage; other-
wise water or cleaning agents may penetrate the system. This can result in malfunctions
or damage to the electrical components.
•
Before operating the bed again, ensure that there is no residual moisture on the electrical
contacts by drying or blowing on the power plug.
26
Part B: Operator and Technical Personnel
Содержание Dali
Страница 2: ...Part A General Information ...
Страница 4: ......
Страница 6: ...II ...
Страница 30: ...Part B Operator and Technical Per sonnel ...
Страница 32: ......
Страница 36: ...IV ...
Страница 88: ...Part C Care staff and residents ...
Страница 90: ......
Страница 129: ...Cleaning and Disinfection Part C Care staff and residents 37 ...
Страница 130: ...Cleaning and Disinfection 38 Part C Care staff and residents ...
Страница 131: ...Cleaning and Disinfection Part C Care staff and residents 39 ...